background image

 

Evaluation Board User Guide 

ADXRS450 - UG 

One

 Technology Way 

 P.O. Box 9106 

 Norwood, MA 02062-9106, U.S.A. 

 Tel: 781.329.4700 

 Fax: 781.461.3113 

 www.analog.com 

 

ADXRS450 Sensor Evaluation System 

 

 

Please see the last page for an important warning and disclaimers.

 

Rev. 0 | Page 1 of 17 

FEATURES 

Flexible inertial sensor evaluation platform 

-

 

Single main board operates with interchangeable 
satellite boards 

Separates DUT from controller for accurate environmental 

testing 

500Hz data rate SPI interface  
Continuous stream to file data recording 
Standard USB cable for power and communications 
PC-based graphical user interface (GUI) 
Fast, easy installation 
 

 
 

 

Figure 1. Inertial Sensor Evaluation System 

 
 

GENERAL DESCRIPTION 

The ADXRS450 inertial sensor evaluation system  
is an easy-to-use evaluation tool targeting bench or desktop 
characterization of Analog Devices, Inc., inertial sensor products. 
The system consists of the inertial sensor evaluation board (ISEB), 
or main board, and a satellite board for any Analog Devices  
inertial sensor product. The ISEB connects directly to a PC via 
USB cable, with the USB connection providing both power and 
communications to the board. The ISEB is connected to the 
satellite board through a ribbon cable. This cable allows the 
satellite to be easily manipulated for testing or for being separately 
placed into an environmental chamber for temperature or 
humidity testing. Separating the boards mitigates corruption  
of data due to the temperature and humidity effects of other 
components. 
The ISEB is a universal main board and is intended to be used 
with various satellites of Analog Devices inertial sensors, including 
analog and digital accelerometers, as well as gyroscopes. The 
different products are evaluated by means of separate GUIs that are 
customized for performance and characterization measurements 
relevant to the inertial sensor being evaluated.  
The EVAL-ADXRS450Z-M system contains the ISEB and the 
EVAL-ADXRS450Z-S satellite. Also included is a USB A to  
Mini-B cable to connect the ISEB to a PC and an 18-inch,  
20-pin ribbon cable to connect the ISEB to the satellite. A CD  

is included with the necessary drivers and installers to use the 
system and to quickly begin evaluating the ADXRS450. 

INSTALLATION OVERVIEW 

The user must complete the following steps, in the order listed, 
in order to be successfully up and running with the ADXRS450 
evaluation system.   

1.

 

Download all CD-ROM files to the PC Hard Drive 

2.

 

Install the USB drivers for the Inertial Sensor 
Evaluation System (ISEB) 

3.

 

Connect the ISEB hardware to the PC 

4.

 

Install the latest Firmware Revision into the ISEB 
hardware (included on the CD-ROM) 

5.

 

Install the ADXRS450 Evaluation System GUI 

6.

 

Configure the ISEB hardware 

7.

 

Launch the ADXRS450 Evaluation System GUI and 
test devices. 

This guide is designed to provide all the detail necessary to both 
install and operate the ADXRS450 Evaluation system. 
 
 

 

Summary of Contents for ADXRS450

Page 1: ...hamber for temperature or humidity testing Separating the boards mitigates corruption of data due to the temperature and humidity effects of other components The ISEB is a universal main board and is...

Page 2: ...ion 4 Updating Repairing the ISEB Firmware 5 Inertial Sensor Evaluation System Graphical User Interface 8 Getting Started 8 View and Record Data 9 Temperature Test 10 Noise and Root Allan Variance 11...

Page 3: ...leted the user can now connect the ISEB main board to the computer via the included USB cable Once connected if the previously installed drivers are not automatically associated with the device then t...

Page 4: ...e Manager a Windows Vista may request that the user allow access to this panel and Administrative privileges may be required b The window shown in XYZ should now open 3 Expand the Ports COM LTP menu i...

Page 5: ...COM port is selected click OK to accept the changes and go back to the main screen Figure 9 Selecting the Correct COM Port for the Downloader When the ISEB is connected the correct firmware is selecte...

Page 6: ...s Figure 6 These directories can be changed by the user if desired however most installations can proceed with the default values Click Next 2 License Agreement Once the installation directories have...

Page 7: ...o generate the device s required high voltage supply For more information on using alternate high voltage supplies refer to the ADXRS450 datasheet as well as the Alternate External High Voltage Supply...

Page 8: ...OM port Refer to the PC s Device Manager under the Ports COM LTP submenu in order to determine which COM port has been assigned to the ADI Inertial Sensor Evaluation System Select this COM port from t...

Page 9: ...software packages However the record functionality can be performed indefinitely if desired by the user Secondary information which is transmitted along with the plotted signals is graphically repres...

Page 10: ...ew and Record panel allowed Data is again streamed to a file to allow for an indefinite length for data recording Before recording the data on this panel if the user wishes to adjust for any null offs...

Page 11: ...t be run at least overnight during which time care should be taken to ensure that the device is in a temperature stable and disturbance free environment Any rotational rate input into the gyroscope ca...

Page 12: ...intended application voltage To record the data from this panel ensure that the Record Bias Run Data box is checked before clicking the View Turn on Bias Stability button The user will be prompted to...

Page 13: ...timings such that this specification is observed then it can be useful to know the behavior of the device from the moment power is applied By sending a series of Sensor Data Requests immediately after...

Page 14: ...by the evaluation GUI The software must perform an initial calibration step in order to accurately calculate the angle of rotation 3 Once the null calibration is performed slowly rotate the satellite...

Page 15: ...d with the high voltage supply the proper current is sourced to the ADXRS450 The supply requirements are described in detail in the ADXRS450 datasheet In short the main requirement for the external hi...

Page 16: ...7 HEADER PINOUT Figure 26 ISEB 20 Pin Header Pinout ESD CAUTION 1 S1 2 3 S2 4 5 S3 6 7 S4 8 CS2 9 VS2 10 VS1 11 12 ST1 13 ST2 14 SCL1 15 SDA1 16 CLK 17 MOSI 18 MISO 19 CS3 20 CS4 P1 3 3V CS2 VS2 VS1 S...

Page 17: ...or use of evaluation boards Information furnished by Analog Devices is believed to be accurate and reliable However no responsibility is assumed by Analog Devices for its use nor for any infringements...