c.
Gentle heating mode: (
)
This mode is suitable for low moisture content, surface casing is easy to form,
easy to burn the sample. From the heat began to rise to the long time
required to set temperature, heating up even slowly. Heating time user
selectable.
d. Press the "MENU" to enter the parameter setting mode, press the "ENTER"
until the heating mode, the icon flashes. Press "UP" or" Down"
choose the mode you need. and press "ENTER" confirm.
3.5.3 Stop mode setting
:(
)
a. Automatic stop mode
:(
)
Heating the sample within the specified time during the weight loss of less
than 2mg will automatically stop when the test. This mode after 15 seconds in
the heating effect.
b. Manual stop mode
:
(
)
Sample heating process is the continuing until the press the "start/stop"
button to end testing.
c. Timing Stop Mode
:(
)
Stop time set by the user, the sample heating process is continued until the
end of the heating time to set the value of the test.
Press the "MENU" button to enter the parameter setting mode, press the
"ENTER" key until the shutdown mode when flashing. Press "UP" or" Down"
to choose the stop mode. and press "ENTER" confirm
,
Parameters are defined,
and go to the next parameter (temperature) setting.
3.5.4
Heating Temperature Setting
The instrument temperature setting range is from 0-199 degrees. Press the
"MENU" to enter the parameter setting mode, press the "ENTER" key until
the
flashing
,
Press "UP" or" Down" to choose the temperature. and
press "ENTER" to go to next setting (time setting)
Icon
will transform other temperature when its heating.
3.5.5
Heating time setting
1.Just choose timing stop mode need set the stop time.
Press the "MENU" to enter the parameter setting mode, press the "ENTER"
key until the
flashing
,
Press "UP" or "Down" to choose the
6