AMTAST MB65 User Manual Download Page 6

heated tube lights, fans turn on, heated symbol in the upper left flashing. While rolling 
on the screen displays the current temperature, has been heating time, moisture 
content of the current sample. 
6. End of the test, heated lamp automatic shut off, the buzzer alarm. Then the sample 
is displayed on the moisture content, heating time, current temperature and the   
parameter setting mode.   

Note: Don't touch the sample plate, the temperature is too high now.

 

7 Press the "Reset" button, the instrument returns to weighing mode. 
 
3.5 Parameters Setting 
3.5.1 Heated parameter storage and recall 
  a. The instrument can store a total of 16 0-F group of heated parameters. Each group 

can set the heating method, stop mode, time and time test parameters. 

  b. In weighing mode, press the "Settings" button to enter the parameter setting mode, 

then store the address number (the number show the highest bit) is flashing. Press 
"down" or "up" key to select the desired address number, the 0 is the factory 
setting,1-F for the user store. 

  c. The memory address number 0 means factory settings, then click "ENTER" button 

to exit the settings back to weighing mode. At this point heated mode is the 
factory defaults. 

  d. When the memory address range for the 1-F any one number, press "ENTER" key, 

and enter the parameter setting after the completion of a parameter (heated 
mode, stop mode, time  mode) setting.  Press the "TARE" key, the address 
number of the test parameters have been stored. 

3.5.2 Heated mode and setting

      

 

a. Standard heated mode

:(

 

 

 

       This mode is the factory setting, suitable for most samples. Heating 

temperature can be set, the sample is heated to a temperature, and keep to 
the test ends automatically. 

    b. Rapid heating mode: (      

 

       This mode is suitable for samples with higher moisture content. You can set 

the heating temperature, heating the temperature will exceeds the set value, 
the future will be reduced to the set temperature and hold until the test ends 
automatically. 

Summary of Contents for MB65

Page 1: ...Moisture Analyzer MB65 MB66 MB67 User Manual Printer is optional AMTAST USA INC ...

Page 2: ...t by youself h Please do not change the instrument s components and any other aspects Maintenance of this equipment by the factory authorized professionals Moisture analyzer use Halogen lamp i Do not put the flammable materials around the instrument j Do touch the halogen lamp when it on working Special attention to the dangers of some sample k Before you test the moisture of dangerous good please...

Page 3: ...uminum pan 100g standard weight 1 piece Product Manual 1 piece Product certification 1 piece Warranty card 1 piece 2 2 Moisture meter placement requirements Operating table Temperature stable good ventilation should be stable dry without strong vibration of the environment without strong magnetic field 2 3 Installation of accessories Place the wind cover aligned holes Place the sample tray holder ...

Page 4: ... side of power cord Please warm up more than 30minutes for keep it stable and accurate III Operation 3 1 Display 3 2 Icon Introduction The percentage of dry weight The percentage of moisture Gram Temperature Time set and heat time Heated mode Standard mode Rapid heating mode Gentle heating mode Stop heated mode Automatic Manual stop mode Timing stop mode Heated state 3 ...

Page 5: ...ET Return to the weighing mode short press calibration balance long press DOWN Reduce UP Add 3 4 First testing 1 First test if not change the temperature and time settings the moisture meter defaults to factory defaults 2 Put empty sample pan on the holder see section 2 3 3 Press the TARE to clean weight 4 Put the sample on the pan Sample weight should be heavy than 1g the general 3 5g 5 Close the...

Page 6: ...he highest bit is flashing Press down or up key to select the desired address number the 0 is the factory setting 1 F for the user store c The memory address number 0 means factory settings then click ENTER button to exit the settings back to weighing mode At this point heated mode is the factory defaults d When the memory address range for the 1 F any one number press ENTER key and enter the para...

Page 7: ...testing c Timing Stop Mode Stop time set by the user the sample heating process is continued until the end of the heating time to set the value of the test Press the MENU button to enter the parameter setting mode press the ENTER key until the shutdown mode when flashing Press UP or Down to choose the stop mode and press ENTER confirm Parameters are defined and go to the next parameter temperature...

Page 8: ...t deficiency can lead to the result of the test a deviation 3 7 Calibration When the instrument error or improper operate Calibration weights can be adjusted with accuracy Press Reset key and does not release until it display cal Then flashing 100 000 put on the calibration weight then it will display until it display 100 000 remove the weight it will display again Waiting it show 0 000 the calibr...

Page 9: ...rder from top to bottom 1 2 3 4 5 1 Aluminum scale weighing tray 2 The triangle scale frame 3 Scale tray brackets 4 Wind cover 5 Stainless steel plate 3 8 the use of the printer and Settings Optional A the realization of the function of moisture meter print must choose my company special printer 7 ...

Page 10: ...e choose to end press confirm button to print the selected the water test report Continue to choose 1 5 then press confirm button to print the selected another water test report 4 Press the reset button to exit the Settings back to the state of weighing Tips 1 Warm up the analyzer 30 minutes and take the first time calibration 2 After the end of a water calibration experiments see 3 6 3 Each time ...

Reviews: