5 — Routine Maintenance
28
11830378_C
AMSCO
®
Rack Return
Operator Manual
Never try to open both doors at the same time.
Never try to open a door manually.
Never interfere with automatic accessory rack movement.
WARNING
PERSONAL INJURY
Never try to intentionally close the door when the safety lock is not properly retracted
during routine maintenance.
Clean equipment with a general purpose cleaner to remove general stains, a stainless-steel stain remover for stubborn
stains, and a stainless-steel polish to keep equipment looking new.
1. Press Command button to open door and put safety lock in place.
NOTE: Safety lock is located on the upper left corner of the door opening.
1
Safety Lock
Figure 5-1. Safety Lock
2. Turn rack return power OFF.
If SCS Conveyor System is present, turn SCS Conveyor System power OFF.
3. Look for any debris or instruments that could have been dropped and left in the chamber. If so, remove it.