3-8
129359-452, Volume 1
Service and Maintenance Procedures
Service Adjustments and Special Features
3.8.2
Undertemperature Recovery Function
NOTE: Resume-time option is only for use in the Laboratory environment. It is
not to be used in a hospital application.
Sterilizer is factory set so that if chamber temperature drops more than 2.0 F*
below set sterilize temperature, and then set temperature is reattained, sterilize
timer will reset to full sterilize time.
Sterilizer can be adjusted so that if chamber temperature drops more than
2.0 F* below set sterilize temperature and then set temperature is reattained,
sterilize timer will resume counting down remaining sterilize time.
3.8.3
Steam Pressure Regulator (Sterilization At Temperatures
Below 250 F)
The steam pressure regulator (HI-LO Valve) is factory set for 250 and 270 F
(121 and 132 C) sterilization cycles. For sterilization temperature control below
250 F (121 C), such as laboratory processes, the LO setting of steam pressure
regulator must be changed by a qualified service technician.
Printout Example
* This value is operator adjustable in the
laboratory mode of operation.