MS322 User's Guide
11
Low Volt AC Stimulator Modality
3. LOW VOLT AC OUTPUT MODE CONTROL
: (SELECTIONS):
Pulsation – Pulsed output (variable from 1 to 80 pulses per second)
Tetanize – Continuous output (set at 80 pulses per second)
Surge – Surged output (set at 80 pulses per second, surge rate
duration variable from 3.75 seconds on, 3.75 seconds off to 0.375
second on, 0.375 second off)
4. PULSE / SURGE RATE CONTROL
:
Pulsation Mode – Regulates the pulsation rate from SLOW (1 pulse
per second) to FAST (maximum of 80 pulses per second)
Surge Mode – Regulates the surge rate from SLOW (3.75 seconds
on, 3.75 seconds off) to FAST (0.375 second on, 0.375 second off)
Note: The Pulse / Surge Rate control has no effect in the Tetanize Mode.
5. OUTPUT RATE INDICATOR LIGHT
: Reflects the output rate for Pulsation,
Tetanize, or Surge Mode.
6. LOW VOLT AC INTENSITY CONTROL
: Regulates the stimulator output
from 0 to MAX. The Intensity control has a unique reset circuit feature
which prevents the practitioner from applying one patient's treatment
settings to another patient. The AMREX Reset Circuit also allows the
continuation or extension of the patient's treatment period if power is
restored within ten (10) seconds.
7. LOW VOLT AC RESET INTENSITY INDICATOR LIGHT
: When power
is restored more than ten seconds after being interrupted or turned
off, or more than ten seconds after the treatment period has been
completed and the bell has sounded, the Low Volt AC Reset Intensity
indicator light will illuminate. The Intensity control must be returned
to the 0/Reset position enabling the audible "click" which will turn
off the Reset Intensity indicator light. NOW the stimulator output can
be activated.
When power is restored within ten (10) seconds after being interrupted or
turned off, or within ten (10) seconds after the treatment period has been
completed and the bell has sounded, the AMREX Reset Circuit will be bypassed
and the Reset Intensity indicator light will not illuminate. This will allow the
continuation or extension of the patient's treatment period.
8. LOW VOLT AC STIMULATOR OUTPUT JACKS
Summary of Contents for MS322
Page 8: ...vi...
Page 12: ...4 MS322 User s Guide Part 1...
Page 16: ...8 MS322 User s Guide Part 2...
Page 20: ...12 MS322 User s Guide Part 3...