3
Note:
1) AC 400.0mV range selection is by RANGE button manually, and is specified
from AC 10mV and up.
2) DC 400.0mV range is designed with 1000M
e
high input impedance for
least current drain in measuring small signals, and can cope better with most
commercially available voltage output transducers and adapters. The non-
zero display reading is normal when the meter inputs are open circuit, which
will not affect actual measurement accuracy. Open input is actually a floating
condition, which is not a zero-volt-input condition. The meter will show zero
or close to zero reading when the inputs are shorted.
e
Resistance, and
Continuity functions
Defaults at
e
. Press SELECT button momentarily to select
Continuity
function which is convenient for checking wiring connections and operation of
switches. A continuous beep tone indicates a complete wire.