EMC GUIDANCE & MANUFACTURER’S DECLARATION
48
D-0115696-C (OM038)
– Otowave 302 Operating Manual
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
– guidance
Voltage
dips,
short interruptions
and
voltage
variations
on
power
supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5
cycle
40% U
T
(60% dip in
U
T
)
for
5
cycles
70% U
T
(30% dip in
U
T
)
for
25
cycles
<5% U
T
(>95% DIP IN
U
T
) FOR 5 SEC
<5% U
T
(>95% dip in
U
T
)
for
0.5
cycle
40% U
T
(60% dip in
U
T
)
for
5
cycles
70% U
T
(30% dip in
U
T
)
for
25
cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
Otowave 302 Tympanometer requires continued
operation during power mains interruptions, it is
recommended that the Otowave 302 Tympanometer
be powered from an uninterruptible power supply or
a battery
Power frequency
(50/60
Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to the application of the test level.