TROUBLESHOOTING
38
D-0115696-C (OM038)
– Otowave 302 Operating Manual
PROBLEM
CAUSE
SOLUTION(S)
AIRFLOW ERROR
Indicator LED c steady light.
Fault with air system and/or pump.
Cannot determine pump direction.
Unknown pump fault. Restart the
unit. If problem persists, contact
your Amplivox service center.
AIRFLOW ERROR
RESTART THE UNIT
Indicator LED c
steady light.
Fault with air system and/or
pump.
Restart the unit. If problem
persists, contact your Amplivox
service center.
WARNING! CALIBRATION
EXPIRED
Indicator LED c
steady light.
The current date is later than the
next calibration date. Check that
the clock is set to the correct date.
If so, arrange for the instrument to
be recalibrated. Tests can still be
performed.
Recalibration
needed
before
further tests are performed.
WARNING! DEVICE
UNCALIBRATED.
Indicator LED c
steady light.
One or more default values
require recalibration before further
tests are performed.
Contact your Amplivox service
center.
WARNING! DEFAULTS
RELOADED.
Indicator LED c
steady light.
Default configuration settings
reloaded.
Default
configuration
settings
reloaded. If the error persists,
contact your Amplivox service
center.
Printing Error
No connection can be
established with the printer
Printer is switched off or not
charged
Connection between printer
and base unit cannot be
established.
Restart the base unit
Restart the printer
Charge printer
Ensure
the
connection
between printer and base
unit is established.
If difficulties resolving fault conditions occur the equipment distributor (or Amplivox if purchased directly) should
be consulted.