background image

 

MAN3069-2 

Page 3 

Ampac Detectors 

 

Cleaning, Maintenance and Functional Test Procedures 

  Place the control panel in ‘test’ mode OR remove the detector from 

the detector base, then re-fit OR reset the panel. 

  Follow steps detailed in 2.1.3 

 

 
 
 
 
 
 
 
 

2.2. 

Testing rate of Rise Detectors  (<90ºC) 

 

2.2.1 Rate of rise heat detectors with fixed tempeatures up to 90°C are tested with the Ampac 

Cordless Heat Detector Tester, part number 204-0017. 

 

2.2.2 Heat detectors with classifications above 90°C (i.e. Series 65 CS/Series 60 Range 2/XP95 

High Temperature Heat Detectors) may require the use of the Ampac mains powered heat 
detector tester, part numbers 204-0005 (240V). 

 

The Ampac ‘Testifire’ unit, part numbers 204-0025, 204-0028 and 204-0029, will test all types of 
heat detector. 

 

2.2.3 Test as per instruction manual supplied with test equipment. 

 

   Insert a charged battery baton into tool and pole. 

   Press the red switch to turn on; LED illuminates green. Slow flash indicates 

normal standby mode. 

   Position tester over detector. 

   Tool will automatically start testing when infra-red beam in cup is broken. 

   Green LED will flash faster. 

   Hold pole in place until the alarm is activated.

*

 

 
 
 
 
 

  When the detector alarms, remove the tester from the detector. 

  The tester will revert to standby mode (slow flashing green LED) and the internal 

fan will cool the heating element for a few seconds. 

 

2.2.4 Additional steps for testing Orbis heat detectors  
 

  For all Orbis heat detectors (even CS class), place the control panel in ‘test’ mode 

OR remove the detector from the detector base then re-fit OR reset the panel. 

  Follow steps detailed in 2.2.3. 

 
 
 

2.2.5 Additional information for testing Discovery heat detectors. 

 

 

Check the control equipment to find which mode the Discovery Heat detector is set to. 
Modes 1–4 will respond using the Series 65/XP95 procedure described in 2.2.3. 

Note: 

The removal and replacement of power to an Orbis detector will place the device into 

‘FasTest’. In ‘FasTest’ the algorithms in the detector are disabled for four minutes. During this 
time the integral red LED will flash once per second. The detector is still able to to detect a fire 
but will respond to one smoke sample in order to speed up testing. This applies to sections 
2.2.4 and 2.3.  

 

*

  If the detector falls to activate, check the heat detector type (i.e. check the temperature 

classification). If the type is correct the device may be faulty. Insert a replacement detector and 
test again. 

 

Summary of Contents for 204-0005

Page 1: ...Ampac Detectors Cleaning Maintenance and Functional Test Procedures MAN3069 2 ...

Page 2: ...est Procedures for Ampac Smoke and Heat Detectors 2 2 1 Testing Ampac smoke detectors 2 2 2 Testing rate of Rise Detectors 90ºC 3 2 3 Testing Multisensor Detector 4 2 4 Testing Discovery CO Detector 5 3 Cleaning Instructions for Optical Smoke and Multisensor Detectors 7 ...

Page 3: ...r with Ampac supplied equipment and may not be used for any other purpose whatsoever 2 The Trader shall disclose Proprietary Information received by Ampac under this Agreement to persons within its organisation only if such persons are legally bound in writing to protect the confidentiality of such Proprietary Information 3 The undertakings and obligations of the Trader under this Agreement shall ...

Page 4: ...pplicator as per instruction manual supplied with the test equipment Place test applicator over detector ensuring seal around smoke detector Push applicator upwards slightly to activate aerosol for a one second burst If the detector has not responded in 10 seconds repeat the previous step The total number of sprays should not exceed five at 10 second intervals If the detector fails to activate che...

Page 5: ...pole in place until the alarm is activated When the detector alarms remove the tester from the detector The tester will revert to standby mode slow flashing green LED and the internal fan will cool the heating element for a few seconds 2 2 4 Additional steps for testing Orbis heat detectors For all Orbis heat detectors even CS class place the control panel in test mode OR remove the detector from ...

Page 6: ... elements can be tested simultaneously using this 2 3 3 For Orbis Multisensor detectors place the control panel in test mode OR remove the detector from the detector base then re fit OR reset the control panel 2 3 4 Follow steps detailed in 2 2 3 2 3 5 For quicker activation times use Testifire with combined smoke and heat test protocol 2 3 6 For XP95 Multisensor detectors follow steps detailed in...

Page 7: ... Place CO test gas inside smoke detector test pole as per instruction manual supplied with test equipment Place applicator over CO detector ensuring seal around CO detector Press upwards slightly to activate aerosol in a one second burst If the detector has not responded in 30 seconds repeat test at 30 second intervals whereby the total number of applications should not exceed five If the detector...

Page 8: ...h Smoke Detector Test Applicator Testifire Cordless Heat Detector Tester Mains Powered Heat Detector Tester Aerosol CO Test Gas Solo C3 used with Smoke Detector Test Applicator Table 1 Summary of functional test procedures Note 1 For Orbis use FastTest Note 2 Discovery detectors may take up to 40 seconds to activate Note 3 Discovery CO detectors may take over 60 seconds to activate dependent on se...

Page 9: ...ssembling Use a commercial vacuum cleaner with a brush attachment to remove exterior dust and dirt and an air pressure canister Detector Duster to blow through the smoke entry points as illustrated in the image below The case can be cleaned with a dampened cloth with a solution of washing up liquid Image A cleaning a detector with a air pressure canister UNCONTROLLED DOCUMENT NOTE Due to Ampac s c...

Reviews: