Amico S-H-300 Operation & Maintenance Manual Download Page 16

16

     Amico Beds Corporation

Preventive Maintenance

WARNING:

 Only facility-authorized personnel should service the Hydraulic Patient Transfer Stretcher  

(S-H-300). Servicing performed by unauthorized personnel could result in injury or equipment damage.

The Hydraulic Patient Transfer Stretcher (S-H-300) requires an effective maintenance program. We recommend that 

you perform annual preventive maintenance (PM) and testing for Joint Commission on Accreditation of Health Care 

Organizations (JCAHO). PM and testing not only meet JCAHO requirements but will help ensure a long, operative life 

for the Hydraulic Patient Transfer Stretcher (S-H-300). PM will minimize downtime due to excessive wear. 

Perform annual preventive maintenance procedures to make sure all Hydraulic Patient Transfer Stretcher (S-H-300) 

components are functioning as originally designed. Pay particular attention to safety features, including but not 

limited to the following:

•  Siderail latching mechanisms
•  Caster braking systems
•  Integrity of sleep surface ticking 
•  Hydraulic functionality
•  Integrity of warning/caution labels
•  Oil leakage from hydraulic piston

Cleaning

BLEACH CLEANING INSTRUCTIONS 

If none of the approved cleaning products provide the disinfectant level required then use a solution of 5.25% Sodium 

Hypochlorite diluted to 1 part bleach to 50 parts water (1000 ppm) (volume to volume). 

Bleach products are corrosive and degrading in nature and may cause damage to your product. If bleach is used, 

measures must be taken to ensure the product is rinsed thoroughly with clean water and thoroughly dried. Failure to 

rinse and dry the product will leave a corrosive residue on the surface possibly causing corrosion. 

•  Failure to follow the above directions may void the product’s warranty.
•  Do not immerse the product in water.
•  Do not attempt to clean or degrease the steel shafts of the hydraulic actuators.
•  Use only approved cleaning products. If bleach is being applied to the product, strictly adhere to the Bleach 

Cleaning Instructions.

•  Do not use a water temperature of more than 50°C or °F. 

All Surfaces 

•  For mattresses please refer to mattress cleaning instructions.

Before Cleaning 

•  Set the stretcher on brake by pressing on the brake pedals to the brake position.

Summary of Contents for S-H-300

Page 1: ...Operation Maintenance Manual Hydraulic Patient Transfer Stretcher S H 300...

Page 2: ...the Reverse Trendelenburg Position 10 To Level the Sleep Surface 10 Head Section Articulation 11 To Raise the Head Section 11 To Lower the Head Section 11 Knee Surface Articulation Crank 11 To Raise t...

Page 3: ...caregivers via pedals located around the center of the base Pedals are located on each side of the stretcher One pedal is used to raise the stretcher and the other pedal is used to either lower or set...

Page 4: ...ectric shock fire and mechanical hazards only in accordance with UL60601 1 CAN CSA C22 2 No 601 1 IEC60601 1 2 38 and IE60601 2 46 CPR control Activates the CPR function Stretcher down control Shows w...

Page 5: ...her Lift Capacity 550 lbs 250 kg Total Weight without Surface 300 lbs max 136 kg max Head Section Inclination maximum 80 5 Knee Section Inclination maximum 30 5 Stretcher Height Range Lowest Position...

Page 6: ...tank holder and patient item storage 3 Transport handles 10 Corner bumpers 4 IV pole mounting socket 11 Brake neutral pedal 5 8 203 mm urethane casters 12 Knee surface articulation crank 6 Fifth wheel...

Page 7: ...recommends that medical personnel determine the proper methods necessary to ensure a patient remains safely in the stretcher TO RAISE A SIDERAIL a Pull the siderail up until it locks into place A cli...

Page 8: ...porated into the stretcher Personal injury or equipment damage could occur Push Handles The push handles located at the head and foot ends of the stretcher are used to move the stretcher a Lift up the...

Page 9: ...ake it easier to turn corners To set the stretcher to neutral set the brake pedal to the middle position When in neutral the brake pedal will be horizontal This allows the stretcher to move in a side...

Page 10: ...until the desired angle is reached TO PUT THE STRETCHER IN THE REVERSE TRENDELENBURG POSITION a Put the stretcher in its raised position see Hi Lo Control on page 10 b Press reverse Trendelenburg con...

Page 11: ...on the head section b Release the bar when the desired position is reached The head section locks in place TO LOWER THE HEAD SECTION a Squeeze the bar under the head section and push down on the head...

Page 12: ...NT IV POLE a Grasp the stored IV pole and raise it from its stored position Make sure it is secure in its support bracket with the sleeve over the support bracket b Grasp the top section of the IV pol...

Page 13: ...ovide a sense of security The use of siderails in any position should be appropriate for the patient s need after assessing any risk factors according to the facility protocols for safe positioning Wh...

Page 14: ...Failure to do so could result in equipment damage CAUTION Do not push or pull the stretcher by IV poles siderails or other equipment Use the push handles located at the head or foot of the stretcher...

Page 15: ...o not exceed 25 psi 172 kPa when spray washing Equipment damage could occur The stretcher can be spray washed as necessary Use a maximum nozzle pressure of 25 psi 172 kPa at a distance of 6 152 mm Spr...

Page 16: ...icking Hydraulic functionality Integrity of warning caution labels Oil leakage from hydraulic piston Cleaning BLEACH CLEANING INSTRUCTIONS If none of the approved cleaning products provide the disinfe...

Page 17: ...ANING INSTRUCTIONS To Wash Wipe with a sponge or soft cloth wetted with warm water containing soap or mild detergent To Rinse After washing rinse cover with warm water to remove residue To Dry Thoroug...

Page 18: ...ording to Amico Beds Corporation specifications used in a normal manner and serviced according to factory recommendations It does not cover failures due to damage which occurs in shipments or failures...

Page 19: ...www amico com 19 Notes...

Page 20: ...H 300 HYDRAULIC PATIENT TRANSFER STRETCHER 03 30 2022 Amico Beds Corporation 85 Fulton Way Richmond Hill ON L4B 2N4 Canada Toll Free Tel 1 877 462 6426 Tel 905 764 0800 Fax 905 764 0862 Email info am...