Safety Tips
FLAMMABILITY
WARNING:
Patients should not be allowed to smoke in the stretcher. Sheets and pillows generally do not
have flame retardant properties. Personal injury or equipment damage could occur.
Reduce the possibility of fires by observing fire prevention rules and regulations. Make sure facility personnel follow
the safety tips in the FDA Public Health Notification: Safety Tips for Preventing Hospital Bed Fires.
SPRAY WASHING
CAUTION:
Do not directly spray the hydraulic cylinders. Equipment damage could occur.
CAUTION:
Do not exceed 25 psi (172 kPa) when spray washing. Equipment damage could occur.
The stretcher can be spray washed as necessary. Use a maximum nozzle pressure of 25 psi (172 kPa) at a distance
of 6" (152 mm). Spray wash comprised of water only should not exceed a temperature of 180°F (82°C). Spray wash
comprised of detergent or solvents (no bleach) should not exceed a temperature of 120°F (50°C). After washing, seal
any exposed chipped steel parts, or oxidized areas with a coat of touch-up paint.