NOTE:
The PSF 16 Channel Power Supply and Filter Pack must be orientated so the
AC power Connections are in the equipment room
. The
product will be labeled SCAN ROOM SIDE.
Diagram 1
- Mounting
Installation:
Shielding Facing
Equipment Room
Shielding Facing
MRI Room
Step 1:
Cut a rough opening in the intended mounting surface measuring
11.25” x 8”. This Step should be performed by a trained RF Vendor/
Installer.
Step 2:
Note whether the Shielding Material of the wall is facing the
Equipment Room or the MRI Room. This will determine how to
secure the PSF 16 Channel Power Supply and Filter Pack in place.
The flange must be mounted against the shielding material with
the supplied gasket in between. Reference
Diagram 1.
Step 3:
Place the provided gasket along the inner edge of the flange so it
is held in place between the flange and the shielding surface.
See
Diagram 2
. Be sure the shielding and power supply plate are clean
before applying the RF Gasket.
Step 4:
Secure the PSF 16 Channel Power Supply and Filter Pack and gasket in place with the supplied screws using the 20 screw holes located
around the perimeter of the flange.
Step 5:
Reference the project specific plans for complete wiring instruction to make the appropriate
Electrical Connections.
Reference
Diagram 4
for instruction on how to open the AC wiring compartment.
Note:
Knockouts are provided. Use an appropriately sized connector clamp to secure the wiring entering and leaving the Filter Pack.
Additional knockouts may be added if necessary.
Note:
A divider will be provided from the factory to separate Class 1 and Class 2 wiring. If using Class 1 wiring, the divider can be removed as
shown in
Diagram 3.
RF GASKET
RF GASKET
RF GASKET
RF GASKET
CLASS 1
WIRING
CLASS 2
WIRING
DIVIDER
Diagram 3
Diagram 4