Amico Pipeline

NOTE: 

Prior to installation, inspect the unit to ensure no tampering or damages were incurred.

1.  Anchor the oxygen inlet station on a finished wall with the appropriate hardware.

2.  Before brazing, wrap wet rags around the tube extension next to the valve body to prevent overheating and 

possible damage to the valve seals. Connect copper tubing to the copper extension using brazing methods and 
materials in accordance with NFPA-99 or CSA Z7396.1.

3.  Test the system for leaks (as per appropriate standards).

To Operate:

1.  Ensure that the isolation valve is closed.

2.  Open the threaded connection.

3.  Connect a pigtail to the oxygen threaded connection.

4.  Check the inlet pressure gauge for supply pressure.

5.  Ensure that the connection is not leaking.

6.  Open the outlet valve.

7.  The “Emergency Oxygen Inlet Station” is now in operation.

To Shut Down:

1.  Ensure that the main bulk oxygen system is working correctly.

2.  Close the isolation valve.

3.  Close the supply pressure to the pigtail.

4.  Open the threaded connection.

5.  Replace the cover plug.

6.  Close the cover and engage the padlock.

Operating Instructions

Installation of Emergency Oxygen Inlet Station

Summary of Contents for M-FILL-OXY-LP

Page 1: ...Operation Maintenance Installation Manual Emergency Oxygen Inlet Station Low Pressure 1 Recessed Mount Model M FILL OXY LP Surface Mount Model M FILL OXY LP S w w w a m i c o c o m...

Page 2: ...edical Facility Responsibility 3 Installation of Emergency Oxygen Inlet Station 4 Operating Instructions 4 Repair and Replacement Parts List 5 Schematic Drawings 6 7 Optional Check and Relief Valves P...

Page 3: ...rovided with a plug to prevent entry of foreign matter into the oxygen system when not in use Amico Emergency Oxygen Inlet Stations will perform in conformity with the descriptions contained in this m...

Page 4: ...em for leaks as per appropriate standards To Operate 1 Ensure that the isolation valve is closed 2 Open the threaded connection 3 Connect a pigtail to the oxygen threaded connection 4 Check the inlet...

Page 5: ...lve seats and O rings 2 V X VLV HAND M Valve handle sleeve for 1 3 V X GAUGE 100 Gauge 100 psi 690 kPa 1 8 NPT 4 M X OXF PLUG LP 1 NPT hex head plug 5 V X BOX PLUG Box plug for oxygen inlet station Bo...

Page 6: ...Connect to pipeline Inlet Oxygen connection 1 NPT Shutoff Valve Provision for Padlock 7 1 4 184 Main Line Shutoff Valve Check Valve 2 required Relief Valve 75 PSIG Turn Down and Screen Main Oxygen Sup...

Page 7: ...XY LP2 Emergency low pressure oxygen inlet station 2 recessed mounted M FILL OXY S Emergency high pressure oxygen inlet station 1 surface mounted M FILL OXY Emergency high pressure oxygen inlet statio...

Page 8: ...co com 85 Fulton Way Richmond Hill Ontario L4B 2N4 Canada Toll Free Tel 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 Email info amico com APE OM INSTAL EOIS LP 2IN 01...

Reviews: