24
Amico Mobility Solutions Corporation
Electro-Magnetic Compliance Data for GoLift Portable450
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.
Emissions test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
√Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable450 and mobile RF communications
equipment should be used no closer to any part
of the [ME EQUIPMENT or ME SYSTEM], including
cables, than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2,5 GHz
Where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
a
,. should be less than the compliance level
in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the [ME
EQUIPMENT or ME SYSTEM] is used exceeds the applicable RF compliance level above, the [ME EQUIPMENT or ME SYSTEM] should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
[ME EQUIPMENT or
ME SYSTEM].
b.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.