www.amico.com
17
Inspection and Maintenance
Prior to using the Amico GoLift Portable450, the inspections should be conducted per the following schedule:
Item
Before Use
Every Month
Annual
GoLift Portable
•
Ensure that endstops are installed
•
Inspect strap for wear or fraying
•
Ensure Batteries are charged
•
Inspect hooks for damage or sharp edges
•
Inspect the wheels in the portable trolley. Replace if damaged.
•
Sling
Check all straps for wear or fraying
•
Inspect the sling for any damage in the fabric
•
Ensure there are no loose threads in the stitching
•
Maintenance by a certified technician
Check the strap and replace only if frayed or damaged
•
Inspect the gearbox for any unusual noises
•
Inspect the hook components for any signs of damage
•
Inspect the gears for any broken or worn teeth
•
** Verify the overspeed cam is operating freely
•
Check emergency stop button
•
** Check emergency lowering device
•
*Annual load test with SWL (maximum safe working load)
•
Ensure the endstops are installed
•
*Reapply grease to the gears
•
* In accordance to the ISO 10535 Standard “Hoists for the transfer of disabled persons – Requirements and test
methods” an inspection should be performed on the GoLift Portable450 at least once a year. This inspection should be
performed by a qualified technician and should include a working load test of one (1) lifting cycle with the maximum
load.
*Re-apply grease to the gears using the Amico certified grease. Can be bought from Amico.
** These two functions must be checked by a qualified technician to ensure the essential performance of the GoLift
Portable450.
!
Do not operate the GoLift Portable450 until any issues discovered during the inspection have been addressed
by a certified technician.