background image

     

www.amico.com    

 17

Cleaning Procedures

Bleach Cleaning Instructions

If none of the approved cleaning products provide the disinfectant level required then use a solution of 5.25% Sodium 

Hypochlorite diluted to 1 part bleach to 50 parts water (1000ppm), (volume to volume).

Bleach products are corrosive and degrading in nature and may cause damage to your product. If bleach is used, 

measures must be taken to ensure the product is rinsed thoroughly with clean water and thoroughly dried. Failure to 

rinse and dry the product will leave a corrosive residue on the surface possibly causing corrosion.

NOTE: 

Failure to follow the above directions may void the product’s warranty.

•  Do not expose the bed to excessive moisture

•  Do not immerse the bed or its components in water

•  Do not attempt to clean or degrease the steel shafts of the actuators

•  Use only approved cleaning products. If bleach is being applied to the product, strictly adhere to the “Bleach 

Cleaning Instructions“.

•  For electrical components, please do not unplug the handset or actuators from the control box for cleaning

•  Do not use a water temperature of more than 50°C or 122°F

NOTE: 

For mattresses, please refer to mattress cleaning instructions.

Before Cleaning

•  Disconnect or unplug the power cord from the main supply

•  Set the bed to BRAKE by pressing on the brake pedals

During Cleaning

•  Do not use abrasive cleaning products

•  Do not use oil-based products

•  Do not use a steam cleaner, washing tunnel, high-pressure spray or hose

To Wash:

•  Wipe with a sponge or soft cloth wetted with warm water containing soap or mild detergent

To Dry:

•  Thoroughly wipe with a dry sponge or soft cloth

To Disinfect:

•  Wash the surface with one of the approved cleaners or refer to the “Bleach Cleaning Instructions“ if bleach 

needs to be used

•  Wipe off residue with water using a soft cloth or sponge

•  Thoroughly dry

Summary of Contents for Apollo MS-SC

Page 1: ...Operation Maintenance Manual Apollo MS SC...

Page 2: ...13 Installation and Removal of the Head Footboards 14 Siderails 14 15 Electrical 15 Preventive Maintenance 16 Cleaning Procedures 17 Cable Schematic 18 Troubleshooting 19 24 Head Siderail Replacement...

Page 3: ...idth 34 86 Caster Size 5 inches diameter 13 cm diameter Bed Lift Capacity Safe Working Load 550 lbs max 250 kg max Total Weight without Surface 450 lbs max 204 kg max Head Section Inclination maximum...

Page 4: ...ification Class 1 Type B Environmental Conditions for Transport and Storage Temperature 40 F to 158 F 40 C to 70 C Relative Humidity 95 non condensing Pressure 50 KPa to 106 KPa Environmental Conditio...

Page 5: ...or protection against fluid ingress and identified as equipment that is protected against spraying and splashing water Ordinary equipment not rated for fluid ingress per IEC60529 This applies to exam...

Page 6: ...30 cm away from the headwall Prior to operating the bed movement controls ensure that all persons or equipment are clear of the bed mechanisms To stop a movement immediately release the button being p...

Page 7: ...for patients in Acute Care and L T A C hospitals This Manual provides instructions required for normal operation of the Apollo MS SC bed from Amico Beds Corporation Prior to operation of this bed it i...

Page 8: ...n Features Item Description Item Description 1 Mattress 6 Base Cover 2 Foldable Siderails 7 Brake Pedal 3 Footboard 8 IV Pole Mounting Socket 4 Corner Roller Bumpers 9 Headboard 5 5 13 cm urethane Cas...

Page 9: ...elenburg Reverse Trendelenburg Bed motions can be controlled via the exterior siderail controls for caregivers and interior siderail controls for patients Optional controls include Footboard controls...

Page 10: ...lenburg Press on this button to place the bed in the desired Trendelenburg position I Bed Flat Press this button to achieve a flat position from any articulated position If a flat patient surface and...

Page 11: ...he patient weight reading To use the Auto Compensate feature Press the button and the LED will begin to blink and will continue blinking until the weight is stable The display will show AUTO during th...

Page 12: ...l Lock Out feature Button K will disable all bed movement controls except for emergency CPR NOTE Lock Out features Buttons J and K will also disable bed movement controls on the Patient Hand Pendant U...

Page 13: ...he CPR Release Lever pull the lever outward with one hand and slowly push the fowler section down with the other hand Unlocking the Bed a To unlock the bed push down on the Green pedal with your foot...

Page 14: ...e bed Ensure that the footboard is properly inserted before use To install the Headboard align the pins with the corresponding plastic guides on the head end support of the bed Insert the headboard by...

Page 15: ...that the battery has a full charge refer to the Batteries section on page 27 The Apollo MS SC features two exterior siderail controls for caregivers and two interior siderail controls for patients The...

Page 16: ...nd testing not only meet JCAHO requirements but will help ensure a long operation life for the Apollo MS SC bed Preventive maintenance will minimize downtime due to excessive wear Perform annual preve...

Page 17: ...he actuators Use only approved cleaning products If bleach is being applied to the product strictly adhere to the Bleach Cleaning Instructions For electrical components please do not unplug the handse...

Page 18: ...B X CBL UBL B X UBL B X UBL B X MJB UBL B X MJB NC BEDEXIT B X CBL 37PIN NC or B X CBL 37PIN NO B X HANDSET 02 B X CABLE MJBSCALE B X CABLE FOOTBOARD B X CABLE SIDERAIL B X CABLE FOOT B X CABLE SIDERA...

Page 19: ...Slot head screwdriver Needle nose pliers Rubber mallet Cable ties Wire cutter Multimeter voltmeter If any cables need to be replaced consult the Cable Schematic and note the corresponding part number...

Page 20: ...ompletely lowered to its Low Height Press Hold the Bed Height UP to raise the bed and confirm that the beeping has stopped Fatal Error On any Bed Control Interface Press Hold the Fowler DOWN and Knee...

Page 21: ...box and the siderails or footboard g To check the electrical connection between the Control Unit and Siderail Controls locate the signal cable as shown in Step 2 of Head Siderail Replacement h If any...

Page 22: ...termine if any of the buttons are being pressed unintentionally b If no buttons are visibly being pressed remove the Footboard and verify if the motion still occurs If motion ceases the Footboard cont...

Page 23: ...the load cells in the wrong direction thus producing a negative weight Check under the Upper Frame of the bed for any obstructions or move the bed away from the obstruction Safe Working Load Exceeded...

Page 24: ...r code reappears contact Amico Beds Customer Service for support Load Cell 2 is Disconnected This occurs if Load Cell 2 is disconnected or has failed Unplug the bed from Mains Power wait 10 seconds th...

Page 25: ...equires replacement Make sure that replacement Head Siderail corresponds to the correct side of the bed 2 Disconnect the signal cable from the main harness To disconnect the cable remove the locking n...

Page 26: ...ations 1 Locate Foot Siderail that requires replacement Make sure that replacement Foot Siderail corresponds to the correct side of the bed 2 Locate the bolts holding the Foot Siderail to the Upper fr...

Page 27: ...e 180 cycles available at 100 discharging between charging 50 discharging depth It appears from the curve that there are 400 cycles available at 50 discharging between charging 30 discharging depth It...

Page 28: ...ected to fall even more WARNING Because of the self discharge you cannot expect the batteries to be fully charged when they leave the factory LINAK recommends that the customer charges the batteries i...

Page 29: ...reach 80 of the battery capacity level We recommend a charging time up to approximately 10 hours which gives 100 capacity At 100 capacity the battery will not discharge as quickly as it would at 80 ca...

Page 30: ...acteristics of the application and the way in which the battery is used WARNING It is not prohibited to use batteries that are not from LINAK s product range as this will result in overloading LINAK s...

Page 31: ...owler Actuator 7 1 B X ACT 01 KNEE LINAK Knee Actuator 8 4 B PCOVER LARMS 001 Lower Hi Lo Arm Cover Top 9 4 B PCOVER LARMS 002 Lower Hi Lo Arm Cover Bottom 10 4 B PCOVER UARMS 001 Upper Hi Lo Arm Cove...

Page 32: ...Corporation Components List THIS PRINT AND INFORMATION HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF AMICO BEDS CORPORATION IT IS NOT TO BE COPIED OR REPRODUCED WITHOUT WRITTEN PERMISSION OF AMICO BED...

Page 33: ...OT Foot End Cover 19 4 B X PGUIDE FOWLER Fowler Plastic Guide 20 2 B X PGUIDE SLIDE Slider Plastic Guide 21 2 B X PBUSHING HILO Hi Lo Bushing 22 2 B X PBUSHING FB Footboard Plastic Insert 23 2 B X PBU...

Page 34: ...ed 7 2 B X BRAKE BUSHCRDL Brake Shaft Bushing Cradle 8 2 B X BRAKE LINK SH Brake Linkage Short 9 2 B X BRAKE LINK LG Brake Linkage Long 10 4 B X PBUSHING BRAKE Brake Shaft Bushing 11 4 B X PBUSHING HI...

Page 35: ...DOWN C Bed Height UP DOWN D Nurse Call E Toggle Underbed Light ON OFF Mattress Please contact Amico Beds for more information WARNING Mattress dimensions must not be smaller than 84 long x 36 wide x 6...

Page 36: ...ROLLER ETS Plastic Bushing B X PBUSHING ETS IV Pole Plastic Insert S X PBUSHING IVPOLE FB Plastic Insert B X PBUSHING FB FB Plastic Insert Slotted B X PBUSHING FB SL Footboard PCB Support B X SUPPCB F...

Page 37: ...R RIGHT Base Cover Left Side B X BASECVR LEFT Caster Arm Bushing B X PBUSHING CASTARM Lower Long Arm Cover Top B PCOVER LARMS 001 Lower Long Arm Cover Bottom B PCOVER LARMS 002 Upper Long Arm Cover Le...

Page 38: ...AD Siderails Foot Right B 4000 SRAIL RFOOT Tente Caster 5 Total Lock Unidirectional B CAS UNI 05B1 Tente Caster 5 Total Lock Directional B CAS DIR 05B1 Brake Pedal S X BRAKE PEDALP Foot Roller B X FOO...

Page 39: ...over Siderail Middle Linkage Covers Footboard Connector B X CABLE FOOTBOARD LINAK DC Actuator Fowler B X ACT 01 FOWLER LINAK DC Actuator Knee B X ACT 01 KNEE LINAK DC Actuator Hi Lo B X ACT 01 HILO NO...

Page 40: ...Cable Red O Ring B X CABLE EXTENSION R Power Supply Cord 120V 126 long cord w 12 pigtail for Grounding H X PCORD 10 NA 37 pin Nurse Call Connector 72 long B X CBL 37PIN 06 Cables WARNING It is NOT re...

Page 41: ...ltage power supply network that supplies buildings used for domestic purposes Harmonic Emission IEC 61000 3 2 Not Applicable Voltage Fluctuation Flicker Emission IEC 61000 3 3 Not Applicable This prod...

Page 42: ...rcial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 U 95 dip in U for 0 5 cycle 40 U 60 dip in U for 5 cycle 70 U 30 dip...

Page 43: ...RF transmitters as determined by an electromagnetic Site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the...

Page 44: ...smitter in Watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all si...

Page 45: ...ng to Amico Beds Corporation specifications used in a normal manner and serviced according to factory recommendations It does not cover failures due to damage which occurs in shipments or failures whi...

Page 46: ...46 Amico Beds Corporation Notas...

Page 47: ...www amico com 47 Notas...

Page 48: ...mico com AB OM APOLLO MS SC 07 29 2019 Amico Beds Corporation 85 Fulton Way Richmond Hill ON L4B 2N4 Canada Toll Free Tel 1 877 462 6426 Tel 905 764 0800 Fax 905 764 0862 Email info amico com www amic...

Reviews: