www.amico.com
9
Preventive Maintenance
WARNING:
Only facility-authorized personnel should service the AC 200 Bed. Servicing performed by unauthorized personnel could
result in personal injury or equipment damage.
The AC 200 Bed requires an effective maintenance program. We recommend that you perform annual preventive
maintenance (PM) and testing for Joint Commission on Accreditation of Health Care Organizations (JCAHO). PM and
testing not only meet JCAHO requirements but will help ensure of a long, operative life for the AC 200 Bed.
Preventive maintenance will minimize downtime due to excessive wear.
Perform annual preventive maintenance procedures to make sure all AC 200 Bed components are functioning as
originally designed. Pay particular attention to safety features, including but not limited to the following:
• Siderail latching mechanisms
• Caster braking systems
• Integrity of sleep surface ticking
• Integrity of warning/cautions labels
• Integrity of the various DC actuators
Cleaning
BLEACH CLEANING INSTRUCTIONS
If none of the approved cleaning products provide the disinfectant level required then use a solution of 5.25% Sodium
Hypochlorite diluted to 1 part bleach to 50 parts water (1000 ppm), (volume to volume).
Bleach products are corrosive and degrading in nature and may cause damage to your product. If bleach is used,
measures must be taken to ensure the product is rinsed thoroughly with clean water and thoroughly dried. Failure to
rinse and dry the product will leave a corrosive residue on the surface possibly causing corrosion.
Failure to follow the above directions may void the product’s warranty.