AMI Analyzer Manual
Expanded Installation Procedure
14
Figure 3. Electrical connections
12. Connect power, analog output and RS485 if desired. Run the power in one conduit, and the
analog output and RS485 in another. You can use flexible conduit for Class 1 Div. 2 areas. See the
NEC.
All the electrical connections are located on a single Phoenix pluggable connector behind a panel on the left
side of the analyzer. Make sure the wires are connected per the markings – Positive to +ve, return to the
ground symbol. This terminal should also be used if required for any shield connections for the output or
RS485 connections. The output connections are made on positions 3 and 4, and the RS485 connections on
5 and 6.
The analog output – which may be either 1-5V or 4-20mA – is connected to the “+” and “-“ symbols (3 and
4). This signal is isolated from ground. If the device to which you are connecting is also isolated from
ground, make sure you ground the “-“ connection at the analyzer. The terminal marked by the ground
symbol may be used for this purpose, in which case the output is no longer isolated. DO NOT GROUND