background image

 

 

270NH3.MAN.REVD.062019 

Page 

8

 of 

17

 

 

 

GENERAL DESCRIPTION, AMMONIA CONVERSION SECTION 

The NH

3

 converter mounts to the right side of a standard 270 series probe enclosure.  The probe enclosure is 

fastened directly to a mounting flange on the stack.  A heat shrink boot on the bottom of the probe enclosure 
provides a sealed input location for all sample lines and electrical wiring for both units. 

Tubing connects one of the heated probe filter outputs to the input on the NH

3

 converter.  This tubing is 

wrapped with a heater to ensure that no condensate forms.  This heater, along with the heater inside the 
converter chamber, is terminated in the probe enclosure, where user power is provided. 

Blowback is typically provided with the probe assembly to flush particulate matter out of the filter.  When the 
accumulator sphere is discharged, small amounts of air will also travel through the converter.  Blowback air 
does not harm the converter but will momentarily disturb both sample outputs.  Clean and dry instrument 
quality air is essential to avoid particulate buildup in the converter chamber.  Typical blowback time intervals 
are user controlled, as frequent as every 15 minutes, but not more` than 24 hours. 

Calibration (cal) gas is injected into the probe chamber, ahead of the filter.  This fulfills most EPA 
requirements for cal gas insertion with respect to proximity of the sample source.  A back-pressure check 
valve, set to approximately 3 psi, is provided in the cal gas injection path to ensure that the gas does not leak 
into the sample path during normal operation.  After the filter, cal gas is drawn through both the converted and 
unconverted sample outputs. 

The converter canister provides a torturous gas path through 316 stainless steel disks and balls, with elevated 
surface temperatures above 1200°F.  The temperature of the chamber is user controlled via monitoring of the 
pre-installed type K thermocouple which is welded to the chamber core.  The oven temperature may be safety 
controlled up to 1500°F; however, the stainless steel surfaces, along with the ammonia gases, oxidize at the 
high end of the range, reducing the service life of the reactor. 

As the sample passes through the reactor, all NO

X

 and NH

3

 are thermally oxidized into Nitric Oxide (NO).  

Reaction of NH

3

 proceeds according to this exothermic reaction: 

4 NH

3

 + 5 O

2

 → 4 NO + 6 H

2

The controlled temperature, gas concentration, and flow rate through the converter should be optimized by 
each user through empirical means.  It is recommended to first test the conversion efficiency at 1250°F.  Most 
applications result in as high as 95% effectiveness at this temperature.  Testing above 1500°F has shown to 
reduce effectiveness in many instances. 

For maximum conversion efficiency and long reaction life (several years), observe the following: 

  Ratio of O

2

 to NH

3

 should be greater than 1.25. 

  Flow through the reactor should be limited to 1 – 1.5 l/m. 

  Do not expose the reactor to ammonia concentrations > 20 ppm. 

  Minimize (NH

3

 ppm * hours * flow rate) for longest reactor life. 

 

 

Summary of Contents for Universal Analyzers 270SF

Page 1: ...5200 Convair Drive Carson City NV 89706 1 775 883 2500 sales universal ametek com www universalanalyzers com Model 270SF Gas Sample Probe with NH3 Convertor Instruction Manual...

Page 2: ...Symbols 4 3 0 Product Identification Configuration 5 4 0 Specifications 6 5 0 Description and Principle of Operation 7 6 0 Installation 9 7 0 Start Up 10 8 0 Shutdown 11 9 0 Maintenance 12 10 0 Troubl...

Page 3: ...cessories immediately on arrival by removing them from the packing and checking for missing articles against the packing list Check the items for any damage which may have occurred during transit and...

Page 4: ...ES EXPOSED SURFACE TEMPERATURE CAN CAUSE BURNS OR PERSONAL INJURY CARE SHOULD BE TAKEN WHEN CONTACT IS REQUIRED CAUTION RISK OF ELECTRICAL SHOCK SYMBOL INDICATES ELECTRICAL SHOCK MAY OCCUR CAUTION SHO...

Page 5: ...e current version of all Model 270SF with Ammonia Convertor product configuration visit the Universal Analyzers website https www universalanalyzers com Navigate to Products Gas Sample Probes Model 27...

Page 6: ...uggested Blowback Timer Period Controlled externally Blowback Solenoid Valve Voltage As specified Maximum NH3 Concentration 20 ppm by volume MATERIALSPECIFICATIONS Enclosure Sample Probe Fiberglass or...

Page 7: ...provided as a means to measure and transmit or record the temperature of the filter A second thermal switch set at 225 F 105 C is provided to be used as an alarm contact if the temperature drops belo...

Page 8: ...lve set to approximately 3 psi is provided in the cal gas injection path to ensure that the gas does not leak into the sample path during normal operation After the filter cal gas is drawn through bot...

Page 9: ...the check valve on the side of the chamber The blowback instrument air tubing connects to the fitting close to the accumulator sphere If an external air line is to be utilized use a bulkhead fitting i...

Page 10: ...r should be installed in the cal gas supply line to ensure that at least 10 more gas is being supplied than what is withdrawn as sample This calculation should include the sum of the unconverted and c...

Page 11: ...ed to temperatures above 300 F and can cause serious burns 4 For further disassembly shut down all heaters in the probe and converter 5 It is recommended to flow instrument air through the converter f...

Page 12: ...r end of the filter Screw the cap back on the filter body The filter replacement procedure is complete NH3 CONVERTER The NH3 Converter requires relatively little maintenance but over a long period of...

Page 13: ...unit Check by measuring for AC voltage on TB1 1 2 Verify by measuring for a closed circuit between TB1 1 4 High oxygen readings low pollutant readings Leak Leaking past the filter cap O ring Remove fi...

Page 14: ...l 270SF with Ammonia Convertor drawings and spare parts visit the Universal Analyzers website https www universalanalyzers com Navigate to Products Gas Sample Probes Model 270SF w NH3 Links to all cur...

Page 15: ...y charge late payment fees at the rate of 1 5 per month or the highest rate permitted by law whichever is less accruing daily B If the financial condition of Buyer is unsatisfactory to Seller Seller m...

Page 16: ...or otherwise THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES WHETHER ORAL WRITTEN EXPRESS IMPLIED OR STATUTORY NO IMPLIED OR STATUTORY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICU...

Page 17: ...persede any prior or contemporaneous representations agreements proposals warranties or understandings oral or written express or implied No waiver modification amendment rescission or other change to...

Reviews: