Calibration
|
5-9
Zero Value: xx
Zero Gas: xx
Zero Drift: xx
The term “zero value” on the display is the value the system read
during the calibrate/verify operation, and the term “zero gas” is the
setpoint value entered using the
Cal Gas Values
option from the
Cali-
brate
key menu.
4. Press
Enter
.
The time when the calibration was completed appear son the display:
End time hr:mn
End date mm/dd/yr
If the date or time has not been set, or a calibration has not been per-
formed, random characters will appear instead of the time and date. If
this occurs, reset the analyzer’s time and date or a calibration is needed.
Verify Data
For verify data, the display shows the verify gas data:
Verify Value: xx
Verify Gas: xx
Verify Diff: xx
The term “verify value” on the display is the value the system read dur-
ing the verify operation, and the term “verify gas” is the setpoint value
entered using the
Configure Verify
option from the
Calibrate
key menu.
The “verify difference” is the difference between the verify gas value that
was read and the verify gas value you entered.
•
Press
Enter
.
The time when the verification was completed appears on the display:
End time hr:mn
End date mm/dd/yr
If the date or time has not been set or a calibration has not been per-
formed, random characters will appear instead of the time and date when
the verification was completed.