4-18 | Thermox CG1000 Oxygen Analyzer
Output Filtering Option
Filter out quick transitions in readings to allow a smoother analog output.
Setting output filtering
1. After selecting an analog output port, from the
Analog Range Key
select
Output Filtering.
2. Enter an output filtering value between 1 and 100 where XX is the cur-
rent filter value, then press
Enter
.
[100 = FAST 1 = SLOW]
Fltr = XX New #?
Note that 100 equals no filtering, and 1 equals the highest possible
filter value. Note that you should enter 01 to enter 1; 07 to enter 7; 70
for 70, etc.
The default value is 100 (no filtering).
Select Function Option
Select what type of information the analog output port should track. Choic-
es are:
•
Oxygen
•
Cell temperature
•
Thermocouple millivolts (T/C mV)
•
Cell millivolts (cell mV)
Selecting the function
1. After selecting an analog output port, from the
Analog Range
menu,
choose
Select Function
.
2. Select what you want the analog output port to track.
The Series 2000 control unit prompts you with the
Set Current Range
menu option
where you can choose the range of readings that the 0 to 20 mA or 4 to 20 mA analog
output represents (see the
Set Current Range
section in this chapter for details).