Mode 3: Auto-Cal 2 pF.
This mode provides a 2 pF preload; alarm setpoint varies depending on
material and coating deposit changes.
Mode 4: Fixed Cal 2 pF.
This mode provides a 2 pF preload; alarm setpoint is locked to starting
capacitance value recorded at system start-up.
Calibration Mode changes
Mode Selection change must be performed with the sensing
element in air (Material below sensing element).
1. On the RF circuit board (
Figure 2-8
), temporarily remove
the shunt jumper from the “Time Delay Selection Jumper”
and place it on pins 1 & 2 of JP5. The green LED will go out,
and the red LED’s will begin to flash. The number of flashes
indicates which mode the unit is in: 1, 2, 3, or 4.
2. To change modes, press and hold the ReCal button (next
to JP 5). The unit will cycle through the modes: first it will
flash one time – then pause, this indicates mode #1. It will
then flash twice then pause, indicating mode #2, then mode
#3, etc. It will scroll through all 4 modes then start over
again at mode #1. Release the button when it reaches the
desired mode. The LED’s will now flash the number of times
indicating which mode has been selected.
3. Remove the shunt from pins 1 & 2 of JP5 and replace it on
the “Time Delay Selection Jumper” pins from which it was
removed. The unit will remain in the new selected calibration
mode. Put the lid back on the housing securely.
2.9.2
IntelliPoint Calibration Mode Change (Continued)
A full system proof test must be performed when:
- Any system component is changed or replaced by the user.
- Any system component or setting is modified by the user.
A full system proof test must be performed when:
- Any system component is changed or replaced by the user.
- Any system component or setting is modified by the user.
Summary of Contents for SIL IntelliPoint RF S*R*L Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 38: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 32 Section 5...
Page 41: ...Specifications 35 Section 6...
Page 43: ...Control Drawings 37 6 1 FM FMc Control Drawings Continued...
Page 44: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 38 6 1 FM FMc Control Drawings Continued...
Page 45: ...Control Drawings 39 6 1 FM FMc Control Drawings Continued...
Page 46: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 40 6 1 FM FMc Control Drawings Continued...
Page 47: ...Control Drawings 41 6 1 FM FMc Control Drawings Continued...
Page 48: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 42 6 1 FM FMc Control Drawings Continued...
Page 56: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 50 6 2 ATEX Control Drawings...
Page 57: ...Control Drawings 51 6 2 ATEX Control Drawings Continued...
Page 66: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 60 6 4 IECEx Control Drawings...
Page 67: ...Control Drawings 61 6 4 IECEx Control Drawings Continued...
Page 75: ...Control Drawings 69 6 5 Mounting and Wiring for Spark Protector Drawings...
Page 77: ...Control Drawings 71 6 5 Mounting and Wiring for Spark Protector Continued...
Page 78: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 72 6 6 Adding a Padded Capacitor...
Page 79: ...Control Drawings 73 6 6 Adding a Padded Capacitor Continued...
Page 80: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 74 6 6 Adding a Padded Capacitor Continued...
Page 81: ...Control Drawings 75 6 7 Dual Seal Assembly for 700 Series Sensing Elements...
Page 82: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 76 6 7 Dual Seal Assembly Continued...
Page 83: ...Control Drawings 77 6 7 Dual Seal Assembly Continued...
Page 84: ...Appendix A...
Page 88: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual A 4 Appendix B...
Page 90: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual B2 EXIDA Report Results...