9.1 SETUP STEPS BEFORE STARTING MEASUREMENT
After setting up the probes as described in
Setting up probes
section, the probe readings has to be reset to a known initial value before starting the
measurement. Below procedure has to be followed
in the same order as mentioned
before starting the measurement:
Perform a PEAK reset on all the probes - Exit from menu screen and set the display in OPERATE screen mode to display the current
readings. Hold DOWN key for 1sec to perform the PEAK reset on all the probes. Alternatively, the operation can also be performed through a
discrete input. Refer to
Input/Output Menu
section for configuration and pin details of discrete inputs. This step will clear all the previously
tracked PEAK+ or PEAK- or run-out values.
NOTE: It takes 0.5sec to perform a PEAK reset on the probe if it is configured in
“Difference mode”
for tracking MAX/MIN/finding run-out. Wait
for the required time according to the number of sensors configured in
“Difference mode”
.
NOTIFY SCREEN
SetUp Menu Page 1
ENTER
Display Configuration
Probes
Computed Measurement
Measurement Mode
Preset/Limits
Next Menu Exit Menu
Probes Menu – Sensor A
Probe Type
Orbit
Direction
Range
Probe ID
Notify Back Next Menu
Probes Menu – Sensor A
Probe Type
Orbit
Direction
Range
Probe ID
Move the Probe Tip or Select
Cancel to end Notify
Cancel Back Next Menu
Probes Menu – Sensor A
Probe Type
Orbit
Direction
Range
+10.00000
Probe ID
110FC49P05
Resolution
18-bit
Notify Back Next Menu
<< On Completion of
NOTIFY process >>
<< Move to the Next
Probe >>