8.11.1 PEAK MODE
PEAK mode is designed to take up to 99 numbers of PEAK readings as input, analyze the same and display the calculated result values. When
PEAK mode is selected, the unit enters into
“acquisition phase”
and the selected channel reading is displayed in single channel mode. User can set
the current displayed reading as a PEAK reading through “ENTER” key press. The storage counter displayed on the screen gets incremented after a
successful store of the PEAK reading. If an input reading has to be erased, the same can be done by PEAK/TRACK key press. When all the
configured number of PEAK readings is input, the unit enters into
“analysis phase”
and the readings are analyzed and below mentioned calculated
result values are displayed through a result message pop-up window:
Max: Maximum PEAK reading in all the PEAK readings input
Min: Minimum PEAK reading in all the PEAK readings input
Range: (Max – Min)
Avg: Sum of all readings/Number of readings input
When the result message pop-up window is exit, all the current readings are erased and the unit is ready to take next set of input readings.
User can print the stored PEAK readings and the calculated result values according to the PRINT mode settings configured. Refer to
PRINT Options
section for further details.
OPTIONS
Advanced Measurement Mode Menu
Advanced Mode
Channel Nr
No. of Readings
Exit Menu
None, Gauging, Autosense, PEAK mode
1 to 31
2 to 99