Page 11
15020-101 Rev-D
Cleaning
CAUTION:
DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY
PART OF THIS INSTRUMENT AS DAMAGE TO THE UNIT MAY OCCUR.
SEE MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTION.
CAUTION:
US
E OF AMMONIA BASED CLEANERS ON
THE LIQUID CRYSTAL DISPLAY
(LCD) MAY CAUSE DAMAGE TO THE DISPLAY. SEE MAINTENANCE
SECTION FOR DETAILED CLEANING INSTRUCTION.
Clean the exterior surfaces of the PDM as necessary. Do not use alcohol or
other solvents to clean the covers, patient windows or screen, as damage may
occur to the instrument.
1. Apply a small amount of mild soap solution to a clean soft cloth and wipe the
exterior of the PDM to remove any contaminants that may accumulate from
normal use.
2. If necessary, apply a small amount of clean water to a clean, soft cloth to
remove any residual soap solution.
3. Dry the exterior using a clean, soft cloth.
4. Isopropyl alcohol may be used to clean the Forehead and Nose Supports.
Remove the Nose Support before cleaning to prevent contacting the Patient
Windows with alcohol. Simply pull the Nose Support away from the instru-
ment, clean, and reinstall.
Battery Replacement
1. Using your thumbs, push and slide the Battery Compartment covers on
either side of the bottom of the PDM.
2. Remove and discard all 4 batteries.
3. Replace them with new AA batteries following the diagram on the inside of
the Battery Compartment.
4. Replace the Battery Compartment covers.
Troubleshooting
Almost all potential problems, such as a no power, dim display, or dim
illumination, are caused by a low voltage condition. When in doubt, always install
new batteries in the instrument.
Cleaning & Maintenance