2. GETTING STARTED
Figure 13
. digiBASE High Voltage Tab
Figure 14
. digiBASE About Tab
dialog, the stabilizer will stay enabled even if the power is turned off. When the stabilizer is
enabled, the
Center Chan
nel and
Width
cannot be changed.
2.5.5. High Voltage
Figure 13 shows the High Voltage
tab, which allows you to turn the
high voltage on or off; and set and
monitor the voltage.
Enter the detector high voltage in
the
Target
field, click
On
, and
monitor the voltage in the
Actual
field. Click the
Off
button to turn
off the high voltage.
2.5.6. About
This tab (Fig. 14) displays hardware
and firmware information about the
currently selected digiBASE as well
as the data
Acquisition Start Time
and
Sample
description. In addition,
the
Access
field shows whether the
MCB is currently locked with a
password (see the password discus-
sion in the MAESTRO
User’s
Manual
).
Read/Write
indicates
that the MCB is unlocked;
Read
Only means it is locked.