digiBASE
®
PMT Base with Integrated Bias Supply, Preamplifier and Digital Multichannel Analyzer
931003E / 0620
CLEAR_ALL
This command clears all spectrum channels, presets, and ROIs. It is equivalent to the
combination of CLEAR_COUNTERS, CLEAR_DATA, CLEAR_PRESETS, and
CLEAR_ROI commands.
CLEAR_COUNTER
Sets the live time and true time counters to zero.
CLEAR_DATA
Sets the channels of spectral data to zero. The ROI flags and presets are not changed.
CLEAR_PRESETS
The live time and true time presets are all set to zero (disabled).
CLEAR_ROI
[start],[length]
If start and length are not specified, the region-of-interest flags for the channels in the
window of interest (see SET_WINDOW) are cleared. If start and length are specified,
region-of-interest flags for the channels specified by start and length are cleared.
DISABLE_GAIN_STAB
Stops stabilization of the gain peak while data is being acquired and sets the center channel
to zero. The gain stabilization adjustment is held at its current value until either gain
stabilization is reenabled with the ENABLE_GAIN_STABILIZATION command or
reinitialized with INITIALIZE_GAIN_STABILIZATION or SET_GAIN_PEAK. See also
SHOW_GAIN_STABILIZATION.
DISABLE_HV
Turns off the HV enable signal. See ENABLE_HV and SHOW_HV.
DISABLE_ZERO_STABILIZATION
Stops stabilization of the zero peak. The zero stabilization adjustment is held at its current
value until either zero stabilization is reenabled with the
ENABLE_ZERO_STABILIZATION command or reinitialized with INITIALIZE_ZERO_
STABILIZATION or SET_ZERO_PEAK. See also SHOW_ZERO_STABILIZATION.
ENABLE_GAIN_STABILIZATION
Enables the stabilization of the gain peak. See also DISABLE_GAIN_STABILIZATION,
SHOW_GAIN_STABILIZATION, and INITIALIZE_GAIN_STABILIZATION.