The counting function of the entire module can be

disabled by holding the Enable input below +1.5 V

using an external signal source. This condition

also turns Off the Gate LED. Open circuit or >+3 V

at the Enable input allows the instrument to count,

if the Count mode has been activated. The Interval

output of another ORTEC timer can perform this

function to synchronize the Model 994 counting

with the other timer. The Interval outputs on all

ORTEC timers provide nom5 V when

counting and <+0.5 V when counting is inhibited.
Independent gating of the A and B Counter inputs

can be achieved with the Gate A and Gate B

inputs on the rear panel. The interface connector

for the RS-232-C option is located on the rear

panel. Each counter has a rear-panel output

dedicated to signaling overflows. Counting these

overflows on another counter extends the counting

capacity of the Model 994. 
The Model 994 derives its power from the ±12 V

and +6 V supplies in a standard NIM bin with

power supply. For bins that do not contain a +6 V

supply, an In6 V Supply option is available.

This option is field-installable and derives its

power from the 117 V AC lines in the bin. 

Specifications 

PERFORMANCE

COUNT CAPACITY

8 decades for counts ranging from

0 to 99,999,999 in each of 2 counters.

MAXIMUM COUNTING RATE

100 MHz for negative

inputs, 25 MHz for positive inputs.

TIME BASE

10-MHz clock with minimum preset or

displayed intervals of 0.01 seconds or 0.01 minutes.

Synchronizing error is nominally 100 ns. Also accepts an

external input from the Counter A input (In A) when the

Ext (External) mode is selected.

TIME BASE INACCURACY

±0.0025% over the 0 to

50°C operating temperature range.

PRESET TIME/COUNTS  

The module stops counting

when the preset value MN X 10

P

is reached on the blind

preset register. M and N are digits ranging from 0 to 6.

With the 0.01-second time base, preset times from 0.01

to 990,000 seconds can be used. Preset times from 0.01

to 990,000 minutes are available using the 0.01-minute

time base. In the Ext time base mode preset counts in

the range of 1 to 99,000,000 can be used. 

POSITIVE INPUT DISCRIMINATOR  

Threshold variable

from +100 mV to +9.5 V with a 25-turn trimpot.

PULSE PAIR RESOLUTION 

<10 ns for negative inputs;

<40 ns for positive inputs. 

INDICATORS

COUNTER DISPLAY

8-digit, 7-segment LED display

with leading zero suppression. When displaying time,

2 digits to the right of a decimal point are included.

OVERFLOW INDICATORS

LED indicators labeled

OVFL A and OVFL B illuminate when the corresponding

A or B Counter exceeds its capacity of 8 decades. The

indicator remains on until a reset is generated.

M, N, AND P INDICATORS

3 LED indicators aid in the

selection of the preset value. When the Preset display

function is activated, the Select push-button selects

which of the 3 LEDs is illuminated. When one of these

LEDs is On, that digit of the preset value can be

incremented using the Advance push-button.

DISPLAY

3 LEDs labeled A, B, and Preset indicate the

information being displayed in the counter display.

Counter A, Counter B, or the Preset value may be

displayed by repeatedly pressing the Display push-

button until the desired LED is illuminated.

TIME BASE

3 LEDs indicate the selected time base

source. By repeatedly pressing the Time Base push-

button, 0.01 Sec, 0.01 Min, or the Ext mode can be

chosen.

GATE 

A single LED indicates that the entire instrument

is enabled to count. For the Gate LED to be illuminated,

the module must be placed in the Count mode (either

manually or via the interface option), the Enable input

must be above +3 V, and the preset stop condition must

not have been reached.

REMOTE 

A single LED labeled REM indicates that the

Model 994 is under computer control, and all front-panel

controls are disabled. This mode is set by the

ENABLE_REMOTE command. 

CONTROLS

DISPLAY

Push-button selects the contents of Counter A

or B, or the Preset value for presentation in the 8-decade

display. Repeatedly pushing the button cycles the

selection through the three choices as indicated by the

A, B, and Preset LEDs.

SELECT 

Push-button chooses the M, N, or P digit in the

display of the preset value. Pushing the button advances

the selection through the three choices as indicated by

the illuminated LED. The Select push-button operates

only if the Preset mode has been selected by the Display

push button. 

ADVANCE

Push-button increments the preset digit

selected by the Select push-button once each time the

Advance button is depressed. The M and N digit ranges

are both 0 to 9. The P digit range is from 0 to 6. The

Advance push-button operates only if the Preset mode

has been selected by the Display push-button.

TIME BASE

Each push on this button advances the

selection one step through the three time base choices

of 0.01 Sec, 0.01 Min, and Ext to determine the time

base source for the preset register.

STOP

This push-button stops all sections of the

instrument from counting.

RESET 

Depressing this button resets both counters to

zero counts and turns Off both overflow indicators. It also

clears any counts accumulated in the blind preset

counter, but does not change the selected preset value.

When power is turned On to the Module, a Reset is

automatically generated.

COUNT

Pushing this button enables the counting

condition for the entire instrument, providing the Enable

input is not held below +1.5 V and the preset value has

not been reached.

THRESH ADJUST (A and B)

Front-panel mounted,

25-turn trimpots to adjust the positive input thresholds for

Counters A and B. The range is from +100 mV to +9.5 V.

Adjacent test points provide the TTL logic signal outputs

from the discriminators to facilitate adjustment using an

oscilloscope.

DWELL

A one-turn potentiometer on the front panel with

an On/Off switch at the fully counterclockwise position.

Adjusts the display dwell time over the nominal range

from 1 to 10 seconds. When the instrument is in the

Recycle mode, dwell time occurs after the preset value

has been reached. Turning the switch Off at the fully

counterclockwise position selects the Single Cycle mode.

If the print loop option is used, the Dwell control is

disabled when the print loop controller is active and

controlling the dwell time.

994

Dual Counter and Timer

2