INPUT POLARITY JUMPERS

Two jumpers located on

the printed wiring board (PWB) separately select the

desired input polarities for inputs In A and In B. P =

positive, N = negative.

A COUNTER/TIMER JUMPER  

Two-position jumper

located on the PWB. In the Counter position, Counter A

always counts and displays the events connected to In

A. When set to the Timer position, Counter A counts and

displays the time if either the 0.01-Sec or the 0.01-Min

time base is selected. If the Ext time base is selected,

Counter A will count and display the events from In A.

B COUNTER/TIMER JUMPER

Two-position jumper

located on the PWB. In the Counter position, Counter B

always counts and displays the events from In B. In the

Timer position with the Ext time base selected, Counter

B counts and displays the time in 0.01-second intervals.

With either a 0.01-Sec or 0.01-Min time base selected,

Counter B counts and displays the events from In B.

GATE A (LIVE TIME/NORMAL) JUMPER 

Two-position

jumper mounted on the PWB. In the Normal position, the

signals from the rear-panel Gate A connector gate the

events from the In A connector. In the Live Time position,

the signals from the Gate A connector gate the 10-MHz

clock to form a live-time clock.

1 CYCLE/RECYCLE

Selection of either the 1 Cycle or

the Recycle mode can be made via an 8-pin DIP switch

on the RS-232-C interface board. The Recycle mode can

be used when the computer is able to respond with a

data transfer when the Model 994 reaches the preset

value. Upon reaching preset, the Model 994 latches its

data into a buffer, resets the counters, and starts the

next counting interval. This process takes ~50 µs. The

computer reads the data in the buffer before the next

counting interval ends. In the 1 Cycle mode, the Model

994 simply stops counting and waits for further

commands when the preset value is reached.

INPUTS

IN A 

Use of this input is affected by the A Counter/Timer

Jumper.

Positive Input  

Front-panel BNC connector for Counter

A accepts positive unipolar signals; minimum width

above threshold, 20 ns at a 50% duty cycle. The

threshold is adjustable from +100 mV to +9.5 V via a

front-panel 25-turn trimpot. Z

in

= 1000 

to ground;

dc-coupled.

Negative Input  

Changing the Input Polarity Jumper

position on the counter board permits selection of the

NIM-standard fast-negative logic input which is designed

to accept –600 to –1800 mV pulses with a fixed

discriminator threshold of –250 mV. Z

in

= 50 

; dc-

coupled. Minimum pulse width above threshold is 4 ns.

IN B 

Identical to In A except that it feeds Counter B. Use

of this input is affected by the B Counter/Timer Jumper.

ENABLE

Front-panel BNC input connector accepts

NIM-standard, slow-positive logic pulses to control the

counting condition of the entire module. A level of >+3 V

or open circuit allows counting provided the instrument is

in the Count mode and has not reached the preset

value; <+1.5 V inhibits counting. The driving source

must be capable of sinking 5 mA of positive current

during inhibit; input protected to +25 V.

GATE A

Rear-panel BNC input connector is identical to

the Gate B input with the following exception. With the

Gate A jumper on the PWB set to the Normal position,

the Gate A input controls counting of the In A events in

Counter A. By moving the PWB Gate A jumper to the

Live Time position, the Gate A input also controls the

10-MHz clock to form a live-time clock with a 100-ns

resolution. A level >+3 V or an open circuit allows

counting of the clock. A level <+1.5 V is used to inhibit

counting of the clock during dead-time intervals.

GATE B 

Rear-panel BNC connector accepts NIM-

standard, slow-positive logic signals to control the

counting in Counter B. A level >+3 V or open circuit

allows counting; <+1.5 V inhibits counting; input

protected to +25 V. The driving source must be capable

of sinking 5 mA of positive current during inhibit.  

OUTPUTS

INTERVAL

Front-panel output BNC connector furnishes

a positive level during the counting interval. The level is

nom5 V when counting is enabled and <+0.5 V

when counting is disabled. Z

o

~30 

.

OVFL A 

Rear-panel output BNC connector provides a

NIM-standard, slow-positive logic signal each time

Counter A overflows its 8-decade capacity. The signal

has a nominal amplitude of +5 V; width ~20 µs.

OVFL B  

Rear-panel output identical to OVFLA except it

monitors overflows from Counter B.

INTERFACE

SERIAL

When the RS-232-C option board is plugged in,

it furnishes a rear-panel, 25-pin, male, D connector

containing all signals for standard RS-232-C

communications. It also contains connections for 20-mA

current loop communications. The field-installable

RS-232-C option provides computer control of the

following functions: Count, Stop, Reset, Remote, setting

the preset value, selecting the displayed counter, and

selecting the desired time base. In the Remote mode,

the computer can disable all front-panel controls.

Computer readout includes: A and B counts, the preset

value, and which counter is being displayed.