ORTEC MODEL 4002P

PORTABLE POWER SUPPLY

1.  DESCRIPTION

1.1.  GENERAL

The  ORTEC  Model  4002P  Portable  Power  Supply 
is ideally suited for providing power  to  preamplifiers,
or  to  a  NIM  module  in  a  remot e  location.  It  is  an
effective solution when the additional power, size, or
cost of a standard NIM bin is inconvenient.

DC  power  is  provided  by  the  Model  4002P  for  four
preamplifiers  through  standard  9-pin  connectors  on
the  rear  panel.  These  connectors  deliver  ±12  and
±24 V dc, and are compatible with ORTEC standard
preamplifier power cables,  as  well as those of most
NIM  manufacturers.  The  Model  4002P  can  supply
power  for  up  to  12  preamplifiers  by  adding  an
optional  Model  4002P-C1  Preamp  Power  Fan-Out
Cable to each connector. The fan-out cable converts
a  single  preamp  power  output  connector  to  three
separate preamp power output connectors. 

A standard NIM bin power connector is incorporated 
for  supplying  ±12  V  dc  and  ±24  V  dc  to  a  NIM
module.  Connection  between  the  4002P  and  the
NIM module is made with the ORTEC Model 401-C3
Module Extender Cable, which can be ordered as an
accessory. The output power is sufficient to operate
a four-wide NIM module and four preamplifiers. 

Designed  for  international  use,  the  4002P  Portable
Power  Supply  can accommodate  input  voltages  of 
100,  120,  220,  and  240  V  ac  at  47  to  63  Hz.  A

connector block on the rear panel is used to change
and  display  the  selec t ed  input  voltage.  Also
incorporated  in  this  connector  block  are  a  fuse
holder  and  the  input  power  cord  connector.  The
4002P  uses  an  international  standard  IEC  power
connec tor  to  permit  the  use  of  power  cords  and
plugs that meet local electrical standards. 

The  Model  4002P  has  maximum  output   current
ratings   of  1  A  on  +12  V, 1 A  on 

!

12  V,  0.5  A  on

+24 V, and 0.5 A on 

!

24 V. 

There  are  three  levels  of  protection  against
overload.  A fuse  on  the  ac power connec tor limits
the  input  current .   All  dc  outputs include  a  current
foldbac k   c ircuit  to  limit  the  output  current  to
nominally  150%  of  the  rated  value.  This  feature
provides  short-circuit  and  overload  protection.
Recovery is  automatic  after  removal  of  the overload
condition.  Test  points  are  conveniently  located  on
the  front panel to allow monitoring  the status  of  the
dc outputs. A temperature warning light on the front
panel turns on when the heat sink temperature rises
above 82

E

C. When the temperature of the heat sink

exceeds   95

E

C,  the  power  supply  is  automatically

turned  off.  Recovery  is automatic  when  the  load  is
reduced  and  the  temperature  decreases  t o  a  s afe
value. 

2.  SPECIFICATIONS

2.1.  PERFORMANCE

INPUT  AC  VOLTAGE

      The  primary  voltage

selection  card  and  indicator  located  in  the  AC
POWER  connector  assembly permit operation  with
100  V,   120  V,  220  V,  or  240  V  nominal  input
voltages.  lnput  voltage  ranges  accommodated  on
each setting are:  

Nominal

Voltage (ac)

Regulation

Range (V ac)

100

88–110

120

103–129

220

191–239

240

206–258

FREQUENCY  RANGE     

Operating  range  for  ac

input voltage frequency is 47–63 Hz.

Summary of Contents for ORTEC 4002P

Page 1: ...Model 4002P Portable Power Supply Operating and Service Manual Printed in U S A ORTEC Part No 740310 0815 801 South Illinois Avenue Manual Revision E Oak Ridge Tennessee 37830 United Statesof America...

Page 2: ...that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the f...

Page 3: ...ACCESSORIES 3 2 7 ORDERING INFORMATION 3 3 INSTALLATION 3 3 1 UNPACKING 3 3 2 SELECTING THE MAINS VOLTAGE 4 3 3 REPLACING THE FUSES 4 3 4 LOCATION 4 3 5 COMPATIBILITY 4 4 OPERATING INSTRUCTIONS 5 4 1...

Page 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Page 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Page 6: ...vi...

Page 7: ...display the selected input voltage Also incorporated in this connector block are a fuse holder and the input power cord connector The 4002P uses an international standard IEC power connector to permi...

Page 8: ...t down condition both the TEMP and the POWER lights are turned off Recovery from thermal shut down is automatic once the load is reduced so that the heat sink temperature drops below the maximum opera...

Page 9: ...monitor each of the dc voltages delivered to the rear panel connectors 2 5 ELECTRICAL AND MECHANICAL WEIGHT Net 10 9 kg 24 lb Shipping 13 kg 29 lb DIMENSIONS 21 5 cm W 26 cm D 10 2 cm H 8 4 in W 10 2...

Page 10: ...n the door window and firmly press the drum back into the holder Close the module door You are now ready to reconnect the instrument to the ac power source 3 3 REPLACING THE FUSES The 4002P requires t...

Page 11: ...e is reached an internal cutout in the power supply removes power and neither lamp will be illuminated Test jacks on the panel allow convenient checking of the power supply voltages from the front pan...

Page 12: ...egulator D4 which receives its unregulated voltage from a voltage doubler circuit consisting of C1 C2 D1 and D2 Current flow to the output is monitored by sensing the voltage drop across R11 When the...

Page 13: ...6 3 FACTORY REPAIR This instrument can be returned to the ORTEC factory for service and repair at a nominal cost Our standard procedure for repair ensure the same quality control and checkout that ar...

Page 14: ...8...

Reviews: