Model 4002P

Portable Power Supply

Operating and Service Manual

Printed in U.S.A.

 ORTEC

®

 Part No. 740310

 0815

801 South Illinois Avenue

Manual Revision E

Oak Ridge, Tennessee 37830
United States of America

Summary of Contents for ORTEC 4002P

Page 1: ...Model 4002P Portable Power Supply Operating and Service Manual Printed in U S A ORTEC Part No 740310 0815 801 South Illinois Avenue Manual Revision E Oak Ridge Tennessee 37830 United Statesof America...

Page 2: ...that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the f...

Page 3: ...ACCESSORIES 3 2 7 ORDERING INFORMATION 3 3 INSTALLATION 3 3 1 UNPACKING 3 3 2 SELECTING THE MAINS VOLTAGE 4 3 3 REPLACING THE FUSES 4 3 4 LOCATION 4 3 5 COMPATIBILITY 4 4 OPERATING INSTRUCTIONS 5 4 1...

Page 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Page 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Page 6: ...vi...

Page 7: ...display the selected input voltage Also incorporated in this connector block are a fuse holder and the input power cord connector The 4002P uses an international standard IEC power connector to permi...

Page 8: ...t down condition both the TEMP and the POWER lights are turned off Recovery from thermal shut down is automatic once the load is reduced so that the heat sink temperature drops below the maximum opera...

Page 9: ...monitor each of the dc voltages delivered to the rear panel connectors 2 5 ELECTRICAL AND MECHANICAL WEIGHT Net 10 9 kg 24 lb Shipping 13 kg 29 lb DIMENSIONS 21 5 cm W 26 cm D 10 2 cm H 8 4 in W 10 2...

Page 10: ...n the door window and firmly press the drum back into the holder Close the module door You are now ready to reconnect the instrument to the ac power source 3 3 REPLACING THE FUSES The 4002P requires t...

Page 11: ...e is reached an internal cutout in the power supply removes power and neither lamp will be illuminated Test jacks on the panel allow convenient checking of the power supply voltages from the front pan...

Page 12: ...egulator D4 which receives its unregulated voltage from a voltage doubler circuit consisting of C1 C2 D1 and D2 Current flow to the output is monitored by sensing the voltage drop across R11 When the...

Page 13: ...6 3 FACTORY REPAIR This instrument can be returned to the ORTEC factory for service and repair at a nominal cost Our standard procedure for repair ensure the same quality control and checkout that ar...

Page 14: ...8...