Page 17
User Guide
Signal Processor
LMG GS
NOTE
If using a thermocouple as a linear 4 to 20mA input, then a transmitter
must be used scaled to fit one of the following ranges:
0 to 600°C 30 to 1100°F
0 to 1100°C 30 to 2000°F
0 to 1600°C 30 to 3200°F
2.2.4 Thermometer input/output connection schedule
The thermometer input/output connection schedules are given in Fig. 12 and 13.
Fig. 12 - GST thermometer cable connection schedules
238012
Pin Nº
Thermometer output
1
2
3
4
5
6
7
8
9
10
----
Green
Yellow
White
Brown
Purple
Screen
Red
Blue
Black
Ground plane
Signal ground (0V)
Signal 1
Signal 2
Signal 3
Signal 4
Cable screen
+V Power supply
-V Power supply
0V Power supply
TH
ER
M
O
M
ET
ER
Cable Colour
Function
NOTE
SIGNAL 3 and SIGNAL 4 are not used for GST thermometers.