Page 11
User Guide
Landcal P80P
Calibration Source
9 Eurotherm Temperature Controller Type 3216CC
EUROTHERM
OP1
Output light
Measured temperature
Setpoint temperature
Scroll
Page
Down
Up
Fig. 3 'Eurotherm' Controller - front panel controls
9.1 Introduction
When switched on, the controller lights up, goes through a short test routine,
and then displays the measured temperature and starts to control. The
output light glows or flashes as heating occurs.
In
Level 1
operation, both the setpoint temperature and the actual measured
temperature are displayed.
The
Page
key allows access to the
Level 2
mode of operation. When in
Level 2
, the parameter lists within the controller can be displayed.
The
Scroll
key allows access to the adjustable parameters within the
controller. Most lists and parameters are hidden and cannot be accessed by
the operator even when in
Level 2
mode of operation. These hidden features
contain factory-set parameters which should not be changed.
The
Up
and
Down
keys are used to alter the setpoint temperature in
Level 1
operation and parameter values when in
Level 2
operation.
To enter the
Level 2
mode of operation:
1) Press and hold the
Page
key for 3 seconds.
2) The display will show
Leu 1 Goto
. Release the Page key.
3) Press the
Up
or
Down
button to choose
Leu 2
(Level 2).
4) Press the
Up
or
Down
button to enter the Level 2 access code, which is
9
.