Ametek Land Landcal P80P User Manual Download Page 13

Page 7

User Guide

Landcal P80P

Calibration Source

6  Using the Source

6.1  Introduction

The P80P has been designed to create an enclosure of uniform temperature, 

ideal for the calibration of radiation thermometers.  The cone point of the 

cavity is placed in the area of minimum gradients within the source.  When 

calibrating radiation thermometers the target size requirements of the 

thermometer should, whenever possible, be fulfilled by the cone.  If the 

thermometer views the walls of the cavity results of greater uncertainty will 

be achieved.

6.2  Measuring sensor (Platinum resistance thermometer) - if supplied

Provision has been made to measure the temperature of the target block using 

a Platinum resistance thermometer which can be inserted from the front of the 

source into the cavity.  When placed in the measuring position the junction lies 

in the same plane as the cone point, but is approximately 40mm/1.5in below.  

The temperature of the source as measured by the sensor agrees with the cone 

point radiance temperature to within ±0.5K (1°F).
The output from the sensor should be measured on an indicator or digital 

voltmeter having a resolution of 0.1°C.
This is the recommended way to obtain the true temperature of the target 

cavity.  The temperature indication on the controller should not be used as an 

accurate measurement of target cavity temperature.

6.3  Operation of the source at below ambient temperature

The LANDCAL Blackbody Source type P80P uses Peltier modules to cool 

the source to temperatures as low as -10°C/15°F.  However the modules 

are rated as only achieving a 30°C (55°F) below ambient cooling effect.  

This means that the final cooling temperature achieved and the length 

of time it takes to achieve this temperature will be dependent on the 

ambient temperature.  For example if the source is operated in an area with 

ambient temperature of 30°C the minimum achievable temperature will be 

approximately 0°C.
The length of time taken to reach a required temperature below ambient can 

be reduced by inserting an insulated plug into the front of the cavity.  This 

plug should be removed at least 15 minutes before any readings are taken, to 

allow the source temperature to stabilise.
Condensation may occur inside the cavity when the source is used at 

temperatures below ambient and above 0°C.  This will in no way affect the 

operation or performance of the source.  At temperatures below 0°C this 

condensation will freeze and form a thin layer of ice.  This ice formation will 

not affect the performance of the source.  If prolonged operation (say longer 

than 8 hours continuous) at temperatures below ambient are anticipated, it is 

recommended that the cavity is purged with a dry inert gas such as Nitrogen 

to prevent entry of water vapour into the cavity.

Summary of Contents for Land Landcal P80P

Page 1: ...Landcal P80P Calibration Source Land Instruments International 2012 User Guide Issue 6 September 2012 Publication N 198 031 Language English...

Page 2: ...n It is possible that this equipment is to be installed on or near to machinery or equipment operating at high temperatures and high pressures Suitable protective clothing along with face and eye prot...

Page 3: ...s etc Design and Manufacturing Standards The Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design manufacture and on site servicing of combustion en...

Page 4: ......

Page 5: ...or Platinum resistance thermometer if supplied 7 6 3 Operation of the source at below ambient temperature 7 7 Calibration of Radiation Thermometers 8 7 1 Preparation 8 7 2 Thermometer calibration 8 7...

Page 6: ...Landcal P80P Calibration Source User Guide...

Page 7: ...he source can be supplied The relationship between the indicated temperature on the controller and the radiance temperature as measured by a secondary standard radiation thermometer is reported 2 The...

Page 8: ...hock never expose the elements terminals or other electrical components when the calibration source is connected to the mains supply After completion of a repair replace all safety plates before switc...

Page 9: ...increase the emissivity value The cavity is heated or cooled using Peltier elements The temperature is controlled by a platinum resistance thermometer connected to a 3 term digital controller having...

Page 10: ...g Design 120 cone Inner diameter 50mm 2 0in Internal length 155mm 6 1in Emissivity 0 995 Controller input Resistance thermometer Controller type Eurotherm 3216 with RS 232 serial interface Electrical...

Page 11: ...ule The colour code for the cable is Brown lead Live Blue lead Neutral Green Yellow lead Earth The source may be connected directly to a 5 amp fused plug and socket The wiring diagram of the P80P is g...

Page 12: ...d at any temperature in the range 10 C 15 F to 80 C 175 F Set the controller to the required value as follows 1 When the source is switched on the fluorescent indicator panel displays the measured val...

Page 13: ...hould not be used as an accurate measurement of target cavity temperature 6 3 Operation of the source at below ambient temperature The LANDCAL Blackbody Source type P80P uses Peltier modules to cool t...

Page 14: ...istance between target and thermometer By adjusting the vertical and transverse vernier screws confirm that the holder is sighted correctly 7 2 Thermometer calibration When soaked conditions have been...

Page 15: ...used in conjunction with an indicator have a resolution of 1K 2 F Fixed installation thermometers whose output is measured on a digital voltmeter will have a resolution of between 0 1K 0 2 F and 0 5K...

Page 16: ...the LAND distributors for repair 8 2 Certification To continue to carry out calibration checks which are traceable to National Standards it will be necessary to obtain a Certificate of Calibration De...

Page 17: ...e of operation When in Level 2 the parameter lists within the controller can be displayed The Scroll key allows access to the adjustable parameters within the controller Most lists and parameters are...

Page 18: ...ns to the measured temperature when no key is pressed for 0 5 seconds 9 3 Altering the Ramp Rate 1 Press the Scroll key until SP RAT SetPoint ramp rate is displayed 2 Use the Down or Up key to adjust...

Page 19: ...ring the Communication Address 1 Press the scroll key until Addr address is displayed 2 To alter the value press the Down or Up key The display returns to the measured temperature when no key is press...

Page 20: ...Page 14 Landcal P80P Calibration Source User Guide 10 Spares No spare parts are available for this instrument Contact LAND if you have a specific requirement...

Page 21: ...ration Accessory enables simple line up of Land fixed installation radiation thermometers When the source is stood on the optional carrying case the bench to mid tar get dimension equals that of the o...

Page 22: ...Page 16 Landcal P80P Calibration Source User Guide...

Page 23: ...rs accessories and mountings and special instruments based on UNO Application dedicated processors based on LANDMARK Graphic ABTS S and ABTS U FTS VDT S and VDT U DTT FLT5 A This 36 month warranty is...

Page 24: ...onditions outside the Ambient Temperature Humidity and Vibration rating of the product environmental contamination solvent vapours deposition of airborne contamination cooling liquids of non neutral p...

Reviews: