6. Use the arrow keys to set the 100% point of the desired pressure range,
select DONE SET when finished.
7. Again, use the arrows to set 0% point and select DONE SET when
finished and the %ERROR mode will be ready to use.
Note:
The 0% and 100% point will be saved in non-volatile memory until
they are changed again by the user for the internal sensors, and external
pressure modules. When using an external module the 0% and 100% are
set to low and full scale of the module until the user changes it, or if it was
previously saved.
31