27
8. Press the “NEXT” option to view when the switch closed, and the dead
band.
9.
Press the “NEW TEST” option to clear the data and perform another
test.
10. Press the “DONE” option to end the test and return to the standard
pressure setting.
Example:
[P1] ST will return to [P1].
Important NOTE:
The previous example uses a normally closed switch.
The basic procedure is still the same for a normally open switch, the display
will just read “OPEN” instead of “CLOSE”.