![Ametek JOFRA HPC600 User Manual Download Page 26](http://html1.mh-extra.com/html/ametek/jofra-hpc600/jofra-hpc600_user-manual_2931362026.webp)
24
Figure 6-4
6. Measuring Voltage
To measure voltage use the input terminals in the front of the calibrator.
Select the Volts function on one of the displays. The calibrator can measure
up to 30V.
Note
: The display will indicate “OL” when the measured voltage exceeds
the nominal range of voltage measurement (30 V).
Figure 7
7. Measuring Temperature with an RTD
To measure temperature using an RTD probe you must select the RTD
function on one of the displays. Make sure the proper probe type is
selected. There are 4 probe types supported, P100-385, P100-392, P100-
JIS and CUSTOM. The standard optional probe has a 250 mm insertion
depth with a 4 mm diameter stainless steel sheath.
Note
: The factory default type is PT100-385 so if the HPC600 is being used
with the AMETEK/JOFRA sensor you do not have to set the probe type.
Simply plug the probe into the HPC600 and configure the display to read
temperature.