Caution…
About the use:
•
Do not use the instrument if the fan is out of order.
•
Before cleaning the calibrator, you must switch it off,
allow it to cool down and remove all cables.
About the well, insertion tube and grid plate:
•
The well and the insertion tube must be clean before
use.
•
Do not pour any form of liquids into the well. It might
damage the well or cause a hazard.
•
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
•
The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
•
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on the insertion tube inside the instrument. There is a
risk that the insertion tube may get stuck if this is
allowed to happen.
Note…
The product liability only applies if the instrument is subject
to a manufacturing defect. This liability becomes void if the
user fails to follow the maintenance instructions set out in
this manual or uses unauthorised spare parts.
Summary of Contents for Jofra ETC-125 A
Page 1: ...Reference manual Temperature Calibrator Jofra ETC 125 A 400 A 400 R ...
Page 3: ......
Page 5: ...2019 04 12 123943 08 List of contents User manual English Reference manual English ...
Page 7: ......
Page 8: ...2019 04 12 ON 115 V 1 5 3 4 2 FIG 1 ...
Page 10: ...2019 04 12 C C AUTO STEP ESC MENU ESC MENU FIG 3 ...
Page 11: ...2019 04 12 FIG 4 AUTO STEP AUTO STEP ...
Page 12: ...2019 04 12 FIG 5 ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ...
Page 29: ...2 2019 04 12 123943 08 ...