
123943
08
15
•
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
4.1
Switching off the calibrator
The following routine must be observed before the insertion tube is
removed and the instrument turned off (cf. Fig. 1) :
1.
If the calibrator has been heated up to temperatures above
50°C/122°F, you must wait until the instrument reaches a
temperature below 50°C/122°F before you switch it off.
2.
If the calibrator has reached a temperature below 0°C/32°F,
it should be heated momentarily to a temperature of
100°C/212°F.
3.
Turn off the calibrator using the power control switch (pos.
4).
4.
Remove the insertion tube from the calibrator using the tool
supplied with the instrument.
Summary of Contents for Jofra ETC-125 A
Page 1: ...Reference manual Temperature Calibrator Jofra ETC 125 A 400 A 400 R ...
Page 3: ......
Page 5: ...2019 04 12 123943 08 List of contents User manual English Reference manual English ...
Page 7: ......
Page 8: ...2019 04 12 ON 115 V 1 5 3 4 2 FIG 1 ...
Page 10: ...2019 04 12 C C AUTO STEP ESC MENU ESC MENU FIG 3 ...
Page 11: ...2019 04 12 FIG 4 AUTO STEP AUTO STEP ...
Page 12: ...2019 04 12 FIG 5 ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ...
Page 29: ...2 2019 04 12 123943 08 ...