Quick Start Guide
1
1.1 System Description
The instructions in this quick start guide are for the AMETEK
Drexelbrook Universal IV Cut Monitor Model Water Cut
Monitor for measurement of the percentage of water in oil. Each
AMETEK Drexelbrook Universal IV Cut Monitor system consists
of a two-wire, 4-20mA electronic unit and a 700 series sensing
element. Communication with the device is done by either an
onboard keypad or with a laptop via HART protocol.
1.2 Unpacking
Carefully remove the contents of the carton and check each item
against the packing list before destroying any packing material.
If there is any shortage or damage, report it immediately to the
factory.
1.3 Safety Information
• Installation, start up, and service should only be
performed by personnel trained in explosive atmosphere
installations
• Substitution of components may impair intrinsic safety
• Live maintenance should only be carried out by skilled
personnel trained in explosion protection methods
• Test equipment used to perform live maintenance must
be certified to use in the associated hazardous area
• When installed as an intrinsically safe device per the
agency control drawings, the housing cover may be
safely opened
• Only cables supplied by Drexelbrook should be used
as the use of other cables can result in unstable
performance.
Explosionproof or Flameproof Installations
• No live maintenance is permitted
• Disconnect power to the device and check that the
atmosphere is clear of hazardous substances
STOP
Section 1: Installation