Quick Start Guide
15
2.5 Calibration / Configuration Via Display / Keypad
• All function in the PC software except Save/Print re
accessible via the display/keypad
• To enter the Configuration Menu follow these steps
• Press and hold the ENTER button for approximately 5
seconds
• Use the UP and DOWN buttons to scroll through the
available selections
• Press ENTER to access sub-menu items
• Use UP and DOWN buttons to adjust settings - settings
that can be adjusted will be flashing
• Press ENTER to accept the adjustment or press and
hold ENTER for 5 seconds to exit to previous menu
Menu Function
(display
abbreviation)
"Values
(display abbreviat
ion)"
Description
"Fct 1.00
Water Cut Ranges
(RANGE)"
Select the water cut range for optimal
measurement
0 to 1% water in Light Oil (LIGHT A)
'Light Oil' is defined as oil with API Gravity
less than 25. Heavy Oil is defined as oil with
API Gravity greater than 25.
0 to 5% water in Light Oil (LIGHT B)
0 to 10% water in Light Oil (LIGHT C)
0 to 30% water in Light Oil (LIGHT D)
0 to 50% water in Light Oil (LIGHT E)
0 to 1% water in Heavy Oil (HEAVY F)
0 to 5% water in Heavy Oil (HEAVY G)
0 to 10% water in Heavy Oil (HEAVY H)
0 to 30% water in Heavy Oil (HEAVY I)
0 to 50% water in Heavy Oil (HEAVY J)
0 to 80% water in Heavy Oil (HEAVY M) -
Default
CUSTOM
Custom range requires a custom strapping
table. See Fct 3.00
"Fct 2.00
Calibration (CAL)"
Enter this menu to calibrate the unit
"Fct 2.01
Indicated Calibration Point
(IND CAL)"
% water - 0.0 Default
Enter the water cut reading captured at the
time of taking the sample for calibration
measurement
"Fct 2.02
Actual Calibration Point
(ACT CAL)"
% water - 0.0 Default
Enter the actual water cut reading verified by
another method of water cut measurement
"Fct 3.00
Strapping table
(STRAP)"
Use this strapping table menu to define
a custom range if selected in FCT 1.00.
Otherwise the correct strapping table is
automatically loaded when range is selected in
FCT 1.00. Default values are for 'HEAVY M'
range
"Fct 3.01
Maximum points
(MAX PNT)"
15 (Default)
Enter the total number of points in the
strapping table which is range dependent
"Fct 3.02
Point number index
(INDEX)"
"1..MAX PNT
1 (Default)"
Enter the point number index