Quick Start Guide
17
2.5 Calibration / Configuration via Display / Keypad (Continued)
"Fct 6.01
Restore factory default
(RST FAC)"
NO (Default)
Select YES to restore factory default in which
case all paramaters will be replaced with
factory default setting. Restoring the factory
default will initiate this message on the display
'DEFAULT PARAMS SET' until power is cycled.
"Fct 6.02
Pad Capcitor in PF
(PAD CAP)"
30.0 (Default)
Enter the value of an external capacitor
that must be connected to the unit. Padding
capacitors are used to reduce the sensing
element standing capacitance in order to
improve the measurement resolution
"Fct 6.03
Contrast
(CONTRST)"
0 (Default)
0 is the highest contrast. 20 is the lowest
contrast
"Fct 6.04
Parameter Number
(PAR NUM)"
0
0 to 65535. Contact factory
"Fct 6.05
Parameter Offset
(PAR OFS)"
0
Contact factory
"Fct 6.06
Parameter Value
(PAR VAL)"
44
Contact factory