Functions
4
nVision Operation Manual
Zero
To zero the nVision:
•
Press the
(
zero
)
button for at least 1 second while vented to atmosphere until the dashed lines (- - - - -) appear.
To clear the zero value:
•
Hold the
(
zero
)
button for 5 seconds until the display changes from (- - - - -) to the zeroed value, then to (- - -).
You can adjust the
Zero Limit
at which the
(
zero
)
button will display “- -HI- -” in
CrystalControl.
You can also disable the
(
zero
)
button
entirely, by setting the zero value to a negative number less than -15 psi.
Note:
If you attempt to zero the gauge with more pressure applied than the Zero Limit set in with CrystalControl, the command will be ignored, and “- -HI- -”
will display.
Note:
You can never zero the BARO sensor.
!
WARNING:
This gauge can display zero pressure when connected to a source of pressure! Do not rely on the display indication before disconnecting—
it may not be indicating true pressure. Never disconnect pressure instrumentation without first relieving system pressure!