nVision Operation Manual
Error 1 displayed
Reason
The nVision checks the integrity of internal calibration coefficients every time it’s turned on. If any coefficients have been corrupted in any way, “Error 1” is
displayed.
Solution
Contact factory for instructions on how to restore the memory to the original factory settings.
Error 2 displayed
Reason
The nVision has tried to display a number too large for the display (i.e., more than 6 digits). May be due to an electrical malfunction or numerical error.
Solution
Contact factory for further instructions.
Error 5 or Error 6 displayed
Reason
The nVision pressure module (PM) is exhibiting out of normal operating condition behavior.
Solution
Contact factory for module replacement.
Error 7 displayed
Reason:
The nVision has lost communication with one or both of your modules.
Solution
It may be possible to recover from this condition by removing and reconnecting your modules using the Module Installation Instructions in this manual. This
may happen if you change modules but do not remove and replace the 4AA, USB, or AC Power to the unit to force a complete reset of the product. Once this
is done the module should be recognized properly. If unsuccessful in resolving the issue, contact factory for module replacement.
Date and Time are incorrect
Reason
The nVision has its date and time synchronised to computers located in our factory.
Solution
Connect to CrystalControl on a computer with the correct date and time settings you desire. During the connection process, the nVision
will be synchronized to the local time.