RADIOLOGICAL MONITOR 

MRS 110B-1 

 
 

 

 

 

 
 

AMES d.o.o. Automatic Measuring Systems for the Environment 

Jamova 39, Ljubljana, Slovenia 

: +365 1 365 71 01, Fax: +365 1 365 71 02 

[email protected]

 

http://www.ames.si

 

 

Summary of Contents for MRS 110B-1

Page 1: ...RADIOLOGICAL MONITOR MRS 110B 1 AMES d o o Automatic Measuring Systems for the Environment Jamova 39 Ljubljana Slovenia 365 1 365 71 01 Fax 365 1 365 71 02 info ames si http www ames si...

Page 2: ...Radiological monitor MRS 110B 1 Manual 2 AMES d o o Jamova 39 Ljubljana Slovenia...

Page 3: ...DETECTED RADIATION 6 1 2 DESCRIPTION OF THE MEASURING METHOD 7 2 USE OF MRS 110B 1 8 2 1 TURNING THE INSTRUMENT ON 8 2 2 WORK WITH THE INSTRUMENT 8 2 3 SWITCHING BETWEEN MEASURING UNITS cpm Sv h 9 2...

Page 4: ...Radiological monitor MRS 110B 1 Manual 4 AMES d o o Jamova 39 Ljubljana Slovenia...

Page 5: ...wall cylindrical halogen Geiger Mueller tube with the cross section of approximately 600 square millimetres Since the device measures roentgen gamma and beta radiation only the equivalent dose rate i...

Page 6: ...easurements are together with the selected measuring unit shown on the liquid crystal display Each detected pulse is heard as an audible signal Different sounding reports that the dose rate has exceed...

Page 7: ...a from the memory from the latest sample back running average until one of the two conditions is fulfilled To sum 150 or more pulses To sum the contents of 30 intervals this means pulses for 1 minute...

Page 8: ...he GM tube When there is no elevated background approximately 20 short beeps are heard each minute MRS 110 B 1 showing aprox 20 cpm If we detect an increased frequency of beeps or increased value on L...

Page 9: ...gy nonlinearity of the GM tube must be taken into the consideration 2 4 SETTING THE ALARM LEVEL To set the alarm level press SET twice After this current alarm level is displayed which can be modified...

Page 10: ...accumulator normal 9 V battery can be used in this case charging is not allowed MRS 110 B 1 detects low battery voltage and warn us with displaying LOBAT on the LCD If instrument is running with the...

Page 11: ...und aprox 15 imp min Detection of roentgen gamma and beta radiation Measuring range 0 01 Sv h to 1 Sv h Display LC display numeric and analogue Alarm adjustable from 0 to 9999 cpm Acoustic alarm Pulse...

Page 12: ...operation This warranty is valid under the following conditions that the device was used in accordance with the supplied instructions that the defect has occurred under the normal use and is not a re...