Amersham Pharmacia Biotech INV-907 Instructions Manual Download Page 6

Valve INV-907

Instructions 56-8101-01 Edition AC

6

Reference information

Description

The valve consists of two main parts:

• Housing which

encloses
the motor and
electronics.

• Valve body with

a rotating
central core.

As the channel
plate is turned by
the motor, different ports are connected.

Valve switching is controlled from
UNICORN by reading the actual position of
the channel plate.

The geometry of the valve assures that
the flow path is completely swept so that
solvent or sample ”memory effect” is virtually
non-existing. The valve rotates the shortest
way to the next position (max. 180°).
The material used in the switching parts
ensures both long mechanical and chemical
lifetime.

The valve housing contains no user
replaceable items.

Technical specifications

Operating data

Max Flow rate

100 ml/min

Max Pressure

25 MPa (250 bar, 3600 psi)

Back pressure

<50 kPa at 100 ml/min with water

Leakage

< 0.1 µl/min at 25 MPa

pH stability range

1–13, 1–14 (<1 day exposure)

Viscosity

Max. 5 cP

Switch time

<260 ms between two 
adjacent positions

Operating life time >50 000 cycles, two 

adjacent positions

Environment

+4 to +40 °C
20-95% relative humidity
84-106 kPa (840-1060 mbar)
atmospheric pressure

Physical data

Internal volume
Pos 1 LOAD

Port 1-7

9 µl

Port 2-3

5 µl

Port 4-6

8 µl

Pos 2 INJECT

Port 1-2

7 µl

Port 3-5

6 µl

Port 6-7

9 µl

Pos 3 WASH

Port 2-4

8 µl

Port 5-7

9 µl

Flow channel 
diameter

0.8 mm

Valve principle

Motor controlled valve

Functions

Switching 3 positions controlled 
from UNICORN

Degree of 
protection

IP 43

Wetted materials 

PEEK (polyetheretherketone)

Chemical 
resistance

The wetted parts are resistant to
organic solvents and salt buffers
commonly used in chromatography
of biomolecules, except
100% ethylacetate, 100% hexane
and 100% tetrahydrofuran (THF)

Power requirement

32 V DC ±10% from the system 
pump

Power consumption

Up to 9 W 

UniNet 2 address

0–9

Inlet and outlet tubing

UNF 10-32 2B ”Fingertights” 
for capillary tubing 1/16”
outer diameter

Dimensions, 
H x W x D

135 x 80 x 120 mm

Weight

1.2 kg

EMC Standards

This product meets the 
requirement of the EMC Directive 
89/336/EEC through the 
harmonized standards 
EN 50081-1 (emission) and 
EN 50082-1 (immunity)
Note: The declaration of 
conformity is valid for the 
instrument when it is
•  used in laboratory locations
•  used in the same state as it 

was delivered from Amersham 
Pharmacia Biotech except for 
alterations described in the 
user manual

•  connected to other CE labelled 

Amersham Pharmacia Biotech 
instruments or other products 
as recommended.

3

INV-907

Housing

Valve 
body

56-8101-01_AC.qxd  1998-10-26 11:45  Sida 6 (1,1)

Summary of Contents for INV-907

Page 1: ...the valve 1 Connect the valve with two UniNet cables as a part of the UniNet 2 chain 2 Set the ID switch at the bottom to the required valve number 0 9 The number should correspond to that used in UN...

Page 2: ...3 Operation The valve is controlled from UNICORN When used as an injection valve use the instruction InjectionValve in System Control Manual Flowpath The valve can be set in 3 positions LOAD position...

Page 3: ...op volume Complete filling 2 5 times the sample loop volume Partial Filling Partial filling is used when high recovery is required Partial filling allows the injected volume to be changed without chan...

Page 4: ...pprox 5 times the sample loop volume should be used to flush the loop and ensure that all sample is injected onto the column Storage Overnight The valve can be left filled with a buffer Weekend or Lon...

Page 5: ...socket 2 Check the ID switch on the valve The ID number should correspond to the number set in UNICORN 3 Check the UniNet cable and replace if required The valve is switching to wrong position The val...

Page 6: ...ysical data Internal volume Pos 1 LOAD Port 1 7 9 l Port 2 3 5 l Port 4 6 8 l Pos 2 INJECT Port 1 2 7 l Port 3 5 6 l Port 6 7 9 l Pos 3 WASH Port 2 4 8 l Port 5 7 9 l Flow channel diameter 0 8 mm Valv...

Page 7: ...1 18 1113 81 Superloop 50 ml 1 18 1113 82 Superloop 150 ml 1 18 1023 85 Cable UniNet 0 7 m 1 18 1109 74 Mounting bracket 1 18 1109 11 Teflon tubing i d 1 8 o d 3 16 3 m 18 1112 47 Tubing connector for...

Page 8: ...Biotech SE 751 84 Uppsala Sweden Trademarks KTA and UNICORN are trademarks of Amersham Pharmacia Biotech Limited or its subsidiaries Amersham is a trademark of Nycomed Amersham plc Pharmacia and Drop...

Reviews: