Manual sample loading

The sample loop and Superloop can be loaded
with a syringe. Connect Superloop with the
bottom tubing connected to port 2 and the
upper tubing to port 6. 

For large volumes use a luer
female/1/16” male union
connected to port 3. 

For smaller volumes use an
injection needle together with
the injection fill port as
described below:

Check the fill port nozzle is screwed into
the fill port. 
Note: The thread is left-handed.

1 Loosely thread the fill port screw into

valve port 3.

2 Insert the injection needle (0.7 mm o.d.)

into the injection fill port.

3 Tighten the fill port until the nozzle has

formed a seal around the needle tip. When
the seal is adjusted correctly, it feels as if
you are penetrating a septum at the end of
the injection fill port. The seal should
provide easy insertion and removal of
the needle.

4. Mount the syringe holder on the fill port.

Two techniques can be used for filling
the sample loop; partial or complete filling.

Type of filling

Volume to load

Partial filling

max 50% of the sample loop 
volume

Complete filling

2-5 times the sample loop 
volume

Partial Filling

Partial filling is used when high recovery is
required. Partial filling allows the injected
volume to be changed without changing
the loop and does not waste sample. 

Partial filling is achieved as follows:

1

Set injection valve to position LOAD.

2 Load the syringe with a large volume of

buffer (5x the loop volume).

3

Fill the sample loop carefully with buffer.

4 Set the injection valve to position INJECT

before taking out the syringe.
Note: If the syringe is taken out when
the injection valve is in position LOAD,
self drainage will occur and air will enter
the sample loop.

5 Load the syringe with the required volume

of sample. No more than half (50%) a
loop volume of sample should be loaded
into the loop.

6 Insert the syringe in position 3 on

the injection valve. Set the injection valve
to position LOAD.
Note: Do not load the sample before
the valve is in position LOAD.

7 Gently load the syringe contents into

the loop.

8 Leave the syringe in position. The sample

will be injected onto the column when
the valve is switched to INJECT in
the method. 

Valve INV-907
Instructions 56-8101-01 Edition AC

3

INV-907

3

56-8101-01_AC.qxd  1998-10-26 11:45  Sida 3 (1,1)

Summary of Contents for INV-907

Page 1: ...the valve 1 Connect the valve with two UniNet cables as a part of the UniNet 2 chain 2 Set the ID switch at the bottom to the required valve number 0 9 The number should correspond to that used in UN...

Page 2: ...3 Operation The valve is controlled from UNICORN When used as an injection valve use the instruction InjectionValve in System Control Manual Flowpath The valve can be set in 3 positions LOAD position...

Page 3: ...op volume Complete filling 2 5 times the sample loop volume Partial Filling Partial filling is used when high recovery is required Partial filling allows the injected volume to be changed without chan...

Page 4: ...pprox 5 times the sample loop volume should be used to flush the loop and ensure that all sample is injected onto the column Storage Overnight The valve can be left filled with a buffer Weekend or Lon...

Page 5: ...socket 2 Check the ID switch on the valve The ID number should correspond to the number set in UNICORN 3 Check the UniNet cable and replace if required The valve is switching to wrong position The val...

Page 6: ...ysical data Internal volume Pos 1 LOAD Port 1 7 9 l Port 2 3 5 l Port 4 6 8 l Pos 2 INJECT Port 1 2 7 l Port 3 5 6 l Port 6 7 9 l Pos 3 WASH Port 2 4 8 l Port 5 7 9 l Flow channel diameter 0 8 mm Valv...

Page 7: ...1 18 1113 81 Superloop 50 ml 1 18 1113 82 Superloop 150 ml 1 18 1023 85 Cable UniNet 0 7 m 1 18 1109 74 Mounting bracket 1 18 1109 11 Teflon tubing i d 1 8 o d 3 16 3 m 18 1112 47 Tubing connector for...

Page 8: ...Biotech SE 751 84 Uppsala Sweden Trademarks KTA and UNICORN are trademarks of Amersham Pharmacia Biotech Limited or its subsidiaries Amersham is a trademark of Nycomed Amersham plc Pharmacia and Drop...

Reviews: