RESIN PACK & FILTER EXCHANGE 

Resin Pack Exchange 

1. 

Turn off the  recirc pump and water supply to the Silex system. 

2. 

Open the air vents located in the center of the cover plates. 

3. 

Place a container under the drain valves located on the bottom plates of 
the system.  Open the valves and allow the water to drain into the 
container.  The container may be emptied at a drain. 

4. 

Remove the cover plates by unscrewing the black knobs located at each 
corner of the cover plates. 

5. 

Pull the exhausted resin pack part of the way out of the column and lay it 
over the top of the system. 

NOTE:

  If the drain valve located on the bottom plate is not open, a vacuum will 

be created making it difficult to remove the resin pack. 

6. 

Slowly pull the exhausted pack over the side of the column while sliding it 
out of the column and into the plastic sleeve that it was shipped in. 

7. 

Remove the new resin pack from the shipping box and open the tied end 
of the plastic sleeve.    Stretch the resin pack over your arm to elongate 
the pack.  Feed the pack into the column allowing it to slide out of the 
plastic sleeve.  Save the plastic sleeve for the return shipment of the resin 
pack at the next exchange.  Close the drain valve. 

8. 

Carefully wipe the O-rings and the grooves on the top plate and cover 
plate to remove any debris from the surfaces that must seal together. 

9. 

Replace the cover plates and tighten the knobs.  To ensure proper 
tightening and a good seal, tighten the knobs in opposite corners 
simultaneously.  Repeat for the other two corners. 

10. 

Turn on the water supply to the Silex system and remove trapped air by 
loosening the air vents located in the center of the cover plates.  Tighten 
the air vents when water begins to escape through the vent.  This ensures 
that all of the air has been removed from the system and maximizes the 
life and efficiency of the resin packs. 

11. 

Allow water to run through the system for 1 minute prior to placing the 
Silex system in service. 

Summary of Contents for SILEX 00M20821

Page 1: ...art Number 98 0097 Revision D SILEX OPERATION MAINTENANCE MANUAL MODEL 00M20821 Ultra Pure Deionizer Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton...

Page 2: ...ESISTIVITY CONTROLLER 6 REMINDER FOR ADJUSTING THE SET POINT 6 SETTING SET POINTS ON RESISTIVITY CONTROLLER 6 ADJUSTING RELAY RESISTIVITY TIME DELAY 7 USING THE RESISTIVITY RELAY 8 RESISTIVITY METER C...

Page 3: ...acks are exhausted replace the packs in the Silex system with stand by resin packs and return the exhausted pack to AmeriWater for regeneration The resin will be regenerated and returned the next busi...

Page 4: ...2 SILEX FEATURES...

Page 5: ...and a 115 Volt receptacle 3 It is recommended to install shut off valves on the inlet and outlet sides of each unit to facilitate changing of resin packs 4 Connect the inlet supply hose to the hose ba...

Page 6: ...nlet of the Silex system to adjust incoming pressure The maximum flow for each deionizer is 3 GPM Gallons Per Minute 7 Install the optional pre filter post filter and or monitors if applicable per the...

Page 7: ...eat for the other two corners e Close the drain valve at the bottom of the column f Repeat steps a through e for each of the columns 9 Open the air vents located on the top of the cover plates and tur...

Page 8: ...hms above the low set point to offset the alarm lock Setting Set points on Resistivity Controller 1 Press and hold the SET POINT button on the resistivity controller and press the ADJUST button 2 Rele...

Page 9: ...To change this digit press DELAY button repeatedly until the desired number is achieved 4 Press the ADJUST button to toggle to the tenths digit 5 To change this digit press the DELAY button repeatedly...

Page 10: ...nection location for the relay in the resistivity controller Resistivity Meter Calibration The resistivity meter should never require calibration It is recommended to verify the accuracy of the resist...

Page 11: ...ilize 9 If the displayed reading differs from the rating of the cal module turn the cal trimmer R16 with a small fine screwdriver until the reading agrees 10 Power down the unit and remove the line co...

Page 12: ...esin pack from the shipping box and open the tied end of the plastic sleeve Stretch the resin pack over your arm to elongate the pack Feed the pack into the column allowing it to slide out of the plas...

Page 13: ...sumps and open the filter drain ports on the bottom of the sumps to drain the water out of the filter housing Close the filter drains when water ceases to flow out of them 3 Use the filter wrench supp...

Page 14: ...you send each pack to AmeriWater enclose the regeneration certificate in the box AmeriWater will ship you a freshly regenerated resin pack 3 To order more certificates contact the company that origin...

Page 15: ...l the presence of chlorine can be detected at the outlet of the system 6 Turn off the water supply to the Silex system and allow the system to soak for 15 minutes 7 After the 15 minute soak time turn...

Page 16: ...val 2 micron 2 5 X 10 20 3050 Post Filter Bacteria and Endo Toxin Removal 05 micron 2 5 X 10 20 5054 Post Filter TOC Removal 5 micron carbon block 2 5 X 10 20 5101 Prefilter Carbon 10 micron block 2 5...

Page 17: ...warranty service notice must be given to the manufacturer within 30 days of the discovery of the defect There are no warranties on the SILEX system beyond those specifically described above All implie...

Page 18: ...16 Addendum A...

Page 19: ...17 AWI 80 0104 GRUNDFOS CIRCULATOR PUMP MANUAL MAINTENANCE FREE CIRCULATOR Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Page 20: ...18...

Page 21: ...19 TABLE OF CONTENTS Shipment Inspection 20 Pre Installation Checklist 20 Installation 20 21 Electrical 21 22 Start Up 22 Operation 22 Failure to Operate 22...

Page 22: ...20...

Page 23: ...21...

Page 24: ...22...

Page 25: ...to cause cancer birth defects or other reproductive harm One such chemical is Vinyl Chloride a compound used to produce Polyvinyl Chloride PVC The AmeriWater system you have purchased may contain PVC...

Reviews: