9
6.
Insert the pins of the cal module (see section 9.0 for part #) in the same
position as the cell wires with the label facing up.
7.
Secure the screws to the cal module.
8.
Plug in the device and turn the unit on, allowing the resistivity display to
stabilize.
9.
If the displayed reading differs from the rating of the cal module, turn the
cal trimmer, R16, with a small fine screwdriver until the reading agrees.
10.
Power down the unit and remove the line cord from the power source.
11.
Remove the cal module and re-install the resistivity cell.
12.
Replace the enclosure’s rear cover and close the electrical enclosure.
13.
Plug the device line cord back into the power supply and turn on the
device.
FIGURE 2